• • • Functional Indicators • • • |
Medium to be measured: | Liquid, gas, or vapor |
Range of measurement: | 0 ~ 0.125KPa ~ 40MPa |
Signal output: | 4 ~ 20mA (expressed with a code E) |
| 4 ~ 20mA, overlapped with HART digital signal (expressed with a code S) |
Power supply of energization | 12 ~ 45V DC (24V DC generally) |
Load characteristic: | RL ≤50 (VS – 12) Ω |
Indicating gauge: | Optional pointer type gauge and LCD |
Explosion – proof: | a. Explosion – isolating type d II BT4 |
| b. Intrinsically safe type in II CT6 |
Measuring scale and zero point: | Externally and continuously adjustable |
Positive and negative immigration: | The lower and upper limit of the scale shall not exceed the scale limit after positive and negative immigration. |
| Maximum positive immigration: 500% of the minimum measuring scale |
| Maximum negative immigration: 600% of the minimum measuring scale |
Range of temperature: | The range of operating temperature for the amplifier: -29 ~ +93℃ |
| The measuring element filled with silicon oil: : -40 ~ +104℃ |
| Flange –type transducer filled with high – temperature silicon oil: : +15 ~ +315℃; that filled with general silicon oil: : -40 ~ +150℃ |
Static pressure and overloading pressure: | 4, 10, 25, 32MPa |
Humidity: | Relative humidity (RH) 0 ~ 99% |
Volume intake capacity: | < 0.12cm3 |
Damping (step response): | Continuously adjustable, between 0.2s ~ 1.67s generally when filled with silicon oil. |
Starting time: | 2s, preheating is unnecessary. |
Accuracy: | ± 0.1% FS; ± 0.25% FS; ± 0.5% FS |
Dead zone: | None (≤ 0.1%) |
Stability: | Not exceeding the absolute value of the basic tolerance for the maximum scale within six months. |
Influence of temperature: | Zero position tolerance is ± 0.1% / 55℃, the total tolerance is ± 0.2% / 55℃. |
Influence of vibration: | On the any axial direction, the tolerance is ± 0.05% / g of the upper limit of the measuring range when the frequency of vibration is 200Hz. |
Influence of power supply: | Less than 0.005% / V of the scale output |
Influence of load: | The load has no influence to it if the power supply is stable. |
Influence of installing position: | 0.24KPa zero point tolerance may be generated, as a maximum, but it may be corrective, without influence to the measuring scale. |
| Isolating membrane: 316LSST, Hastelloy alloy C, MONEL, or tantalum. |
| Gas exhaust / liquid discharging valve: 316LSST, Hastelloy alloy C, MONEL |
| Flange and joint: Electroplated carbon steel, 316LSST, Hastelloy alloy C, or MONEL |
| O – ring contacting medium: NBR, fluorubber |
| Liquid filled: Silicon oil or inertia oil |
| Bolt: Electroplated carbon steel |
| Enclosure of electronic parts: Low – copper aluminum alloy |
Pressure – leading connecting part: | The connecting screw hole on the pressurized vessel or chamber is 1 / 4 ~ 18NPT, the connecting screw hole on the pressure – leading joint is 1 / 2 ~ 14NPT. |
Connecting hole of the signal wire: | G1 / 2 |
Weight: | Approximately 3.5kg (excluding accessories) |
Standard accessories: | The flanged joints, gas exhaust valves and liquid discharging valves and one book of instruction manual have been provided for all the variants in the factory before delivery. |
| |