Commercial Freeze Dryer for Sale
KFD Series Medium-size Laboratory lyophilizer
Technical parameters
| Model | KFD-1 | KFD-2 | KFD-5 | KFD-10 | KFD-20 |
| Type | Conductive Type |
| Ice Capacity (Kg/batch) | 15 | 30 | 75 | 150 | 300 |
| Condenser Temperature (℃) | <-45℃ |
| Shelf Temperature (℃) | -30~60℃ |
| Pre-freezing | In Original Place |
| Shelf Area (㎡) | 1 | 2 | 5 | 10 | 20 |
| No. of Shelves | 5+1 | 6+1 | 5+1 | 11+1 | 14+1 |
| Shelf Dimensions ( m) | 0.5*0.4 | 1.0*0.4 | 1.25*0.8 | 1.2*0.8 | 1.42*1.0 |
| Shelf Spacing (mm) | 60 | 60 | 70 | 70 | 70 |
| Medium | Silicon Oil |
| No. of trays | 5pcs | 12pcs | 30pcs | 66pcs | 126pcs |
| Tray Dimensions (mm) | 400*500 | 345*95 | 395*415 | 395*395 | 470*325 |
| Refrigerant | R404A |
| Cooling Method | Air Cooled /Water Cooled |
| Final Vacuum(Pa) | 10Pa |
| Deforesting Method | Hot Water or Steam |
| Total Power (KW) | 9KW | 13KW | 20KW | 40KW | 50KW |
| Installation Area (㎡) | 3 | 6 | 9 | 11 | 15 |
| Chamber Dimension (m) | 1.2*0.6*1.8 | 1.5*1.0*1.8 | 2.5*1.3*2.0 | 3.0*1.3*2.5 | 3.7*2.0*2.5 |
| Weight (kg) | 1500 | 2000 | 2500 | 3800 | 8500 |
Installation confirmation of commercial freeze dryers for sale
1. Confirmation of installation files
2. Confirmation of main material
3. Confirmation of main equipment supporting
4. Confirmation of public supporting works
5. Confirmation of equipment manufacturing quality
6. Confirmation of instrument calibration and certificate of conformity
Operation Confirmation of Commercial Freeze Dryers for Sale
1. Verification of plate temperature uniformity and temperature control ability
2. The cooling rate and low temperature of the board layer
3. Layer heating rate and high temperature
4. Condenser cooling rate and low temperature
5. Pumping rate and ultimate vacuum
6. Vacuum leaking rate
7. CIP in-place cleaning effect verification
8. Verification of online (ozone) sterilization effect
9. Safety performance test
10. Verification of respirator integrity
11. Verification of the tampon system
Performance Confirmation of Commercial Freeze Dryers for Sale
1. Verification of the temperature uniformity and temperature control ability of the board layer
2. Verification of layer cooling rate
3. Verification of the heating rate of the board layer
4. Verification of the cooling rate of the condenser
5. Verification of the pumping rate
6. Verification of large catching of ice
The best installation state of the Freeze Dryer:
1. The machine must not be placed in direct sunlight, so as not to affect the heat transfer effect
2. Try to install the freeze dryer in a well-ventilated place, and there should be no obstacles in front of and behind the ventilation.
3. If there are multiple freeze dryers in a room, the freeze dryers must be arranged horizontally to avoid affecting each other.
4. If there is too much dust in the installation environment of the freeze dryer, it is necessary to set up a filter cover in front of the air inlet that is easy to clean and does not affect ventilation, and clean it frequently so as not to affect the normal operation of the freeze dryer.
5. There should be no heat source near the freeze dryer.
Application
Looking for ideal Commercial Freeze Dryer Hot Sale Manufacturer & supplier ? We have a wide selection at good prices to help you get creative. All the Freeze Dryer for Vegetables are quality guaranteed. We are China Origin Factory of Freeze Drying Machine Factory Price. If you have any question, please feel free to contact us.
Product Categories : Freeze Dryer > Commercial Freeze Dryer