Welcome Customer !

Membership

Dongguan Tianhai Quality Inspection Instrument Co., Ltd.
Free Member ·1YRS
DongGuan,GuangDong
Manufacturer

Main Products: Test equipment

Xwboo>Products
Product Categories

Dongguan Tianhai Quality Inspection Instrument Co., Ltd.Free·1YRS

  • E-mail

    tianhaiyiqi@126.com

  • Phone

    0769-85335569,85351859,85333569

  • Address

    2nd Floor, No. 49, Qilong Road, Xiabian Community, Chang'an Town, Dongguan City, Guangdong Province

Contact Now

TH8145 Halogen acid gas release measuring device

Model
TH8145
Nature of the Manufacturer
Manufacturer
Product Category
Plastic testing equipment
Place of Origin
DongGuan City, GuangDong Province
Overview
TH8145 Halogen acid gas release measuring device
Product Details

Overview:

It is used to determine the total amount of halogen acid gas in low-halogen and halogen-free flame-retardant cables and to measure the acidity of the gas by measuring the PH value and conductivity.
Comply with standards: GB/T17650.1-1998, GB/T17650.2-1998 and IEC60754-1:1994, IEC60754-2:1991 standards; at the same time meet the GB/T19666-2005 "Flame-retardant and fire-resistant wires and cables" standard Table 6 regulations and GA306.1-2007 Article 6.3.4, GA306.2-2007 Article 6.3.4, UL2556-2007 Article 9.10 and 9.11 requirements.

Main technical parameters:
1. Tubular furnace: The effective length and diameter of the heating chamber are 600× ф42mm, equipped with an adjustable electric heating system. Heating power: 220V /10KW
2. Combustion tube:
1. Material: Refractory pipe made of silica, resistant to corrosive gas.
2. Specifications: The inner diameter is 40mm, the length extending from the inlet side of the tube furnace is 60~200mm, and the length extending from the outlet side is 60~100mm, leaving a gap for thermal expansion.
3. Burning boat for measuring the release of halogen acid gas:
1. Material: Use fused quartz tube.
2. Specifications: length 100mm, width 25mm, depth 10mm.
3. Combustion boat feeding method: using bar magnet and platinum wire pulling method.
4. Gas collection: Two selected bottles are used: one of which is equipped with a magnetic stirrer and contains 220 mL of sodium hydroxide solution.
5. Gas supply system:
1. Air flow meter: 20mL~200mL adjustable
2. Compressor (provided by your company)
6. Sampling: 500mg~1000mg
7. Determination of the halogen acid content of the halogen acid gas release measuring device:
1. Analytical balance: accuracy of 0.1mg
2. Digital PH meter: the accuracy is 0.02PH
3. Conductivity meter: Measuring range: (0~1×105)μS/cm.

Similar Product Recommend