HZSTM-4 Stormer Viscometer Application Manual

Model
Nature of the Manufacturer
Manufacturer
Product Category
Oil Viscosity Tester
Place of Origin
City, Province
Overview
HZSTM-4 Stormer Viscometer Application Manual
Product Details

I. General Introduction

This meter suits to measure paint and other with KU value to indicate the coating viscosity. It can display the viscosity KU value of measured samples or the load value related to KU value (unit: g).

It is controlled by micro computer .Dispersing disc driven by special motor with constant speed to 200r/min. The resistance torque of the agitation disc receives during rotating of the sample measured will be transferred by computer to kind of logarithmic function with KU value required by the load for producing 200r/min(Normally to indicate the viscosity of brush coating and roll coating.)The operator just have to put the agitating disc into the sample to the regulated depth, the KU value and related load value can be read directly from the meter.

The meter is designed based on ASTM-D562 and GB9269-88.

II.Tech Parameter

a) Measuring range: 40.2KU~141.0 KU(KU value)

32g~1099g(Load value)

b) Measuring accuracy : 1.5% of full measuring travel

c) Repeatability : 1.5% of full measuring travel

d) Agitating shaft speed :200r/min±1r/min

e) Capacity of vessel : around 500ml

f) Input voltage : 220V 50Hz

g) Power : 10w

h) Overall size: 210mm×180mm×500mm (L×W×H)

i) Weight: 10kg

III.Test environment

j) The meter should be located on firm and level smooth table far from interference of strong airflow, magnetic field, electric pulse and electric field.

k) For the security of operator, the socket of power should be put on the ground reliably.

l) Temp. of laboratory should be 23℃±2℃ and relative moisture ≤ 75℃.

m) Temp. of sample : 23℃±0.2℃