9187sc chlorine dioxide analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
Chlorine dioxide analyzer
Place of Origin
City, Province
Overview
9187sc chlorine dioxide analyzer
Product Details
typical application
Suitable for online monitoring of chlorine dioxide concentration in drinking water and industrial process water disinfection and sterilization processes
 
Instrument characteristics
● Wide measuring range
● Using selective membrane sensor
● Low maintenance: calibration once / 2 months; replacement of electrolyte and membrane / 6 months
● No interference from chlorine, bromine or hydrogen peroxide; O3 is interference, but O3 is not available in the general chlorine dioxide disinfection system
 
Measuring principle
The electrolyte and the permeable membrane separate the electrolytic cell and the water sample, and chlorine dioxide can penetrate the membrane; there is a fixed potential difference between the two sensors, and the current generation is proportional to the concentration of chlorine dioxide;
On the cathode: ClO2 + 4H + + 5e- → -Cl- + 2H2O
On the anode: Cl- + Ag → AgCl + e-
The sensor has a temperature probe, its function is temperature compensation
 
Scan the QR code to read on your phone
Technical index
Measuring range0.010 - 2 ppm (mg/L) (ClO2)
Detection limit10 ppb or 0.01 mg / L ClO2
Accuracy5% or ± 10 ppb ClO2
Response time90% less than 90 seconds
Sample flow rate200~250 mL/min
storage temperature-20~60℃
Operating temperature0~45℃
Sample temperature0~45℃
Correction methodLaboratory comparison
Calibration intervalOnce / 2 months
Maintenance intervalGenerally replace the membrane and electrolyte every six months
Injection connection1/4-in. O.D.
Emission connection1/2-in. I.D.
Protection levelIP-66/NEMA 4X
Instrument size270 x 250 mm

 