5500sc-AMC ammonia_monochloramine analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
An ammonia analyzer
Place of Origin
City, Province
Overview
5500sc-AMC ammonia_monochloramine analyzer
Product Details
Make the chloramine process easier
 
Hach Company 5500sc ammonia / monochloramine analyzer can provide all the information you need in the chloramine process to avoid nitrification, odor and odor, so that you have full confidence in the chloramine process.
 
Technically reliable
 
Hach Company provides you with a very low-maintenance solution with leading technologies such as pressurized reagent delivery system, easy-change reagent bottles, simple fault diagnosis menu, and PROGNOSYS predictive diagnosis system. Without much attention from you, 5500SC can do the work by itself.
 
Strengthen control
 
Continuous online monitoring allows you to understand the plant's chloramine process more accurately and completely, and make data-oriented judgments. The dual colorimeter design allows you to obtain data faster, and is always a step faster than the nitrification process.
 
Easy operation
 
You only need to set up 5500sc to start working. Online monitoring of the chloramine process has never been easier. Only through the status indicator, friendly user interface, reagent bottle color indication and a minimalist program to verify the sampling results, it can be run well.
 
Principle of operation
 
The analyzer measures total ammonia and monochloramine in drinking water to determine the concentration of free ammonia. The improved phenate colorimetric method is used to determine the concentration of monochloramine. Under excess hypochlorite and a suitable pH value, the total ammonia concentration can be measured by the phenate method. The total ammonia minus monochloramine gives the free ammonia concentration.
 
Scan the QR code to read on your phone
Technical index
Range0.01 ~ 2ppm (calculated as N)
Measurement parametersTotal ammonia, monochloramine, free ammonia
Number of sample streams1 or 2, programmable sequence
Accuracy
5 to 40 ° C (41 to 104 ° F), ± 5% or 0.01 ppm (in N)
40 ~ 50 ° C (104 ~ 122 ° F), ± 10% or 0.02ppm (calculated as N), whichever is greater
Repeatability3% or 0.01ppm (calculated as N), whichever is greater
Lower detection limit0.01ppm (in N)
Response timeWithin 5min
Medicament consumptionWhen the circulation time is 5min, 1L / month
Operating temperature range5~45 °C(41~122 °F)
Operating humidity5 ~ 95% non-condensing
Sample pressure0.17~6.0bar(2.5~87psi)
Sample pressure5~50 °C(41~122 °F)
Sample tassel100~1000mL/min
Sampling abilitySampling access capacity
Installation methodWall, panel or table
Import connection6mm outer diameter plastic tube pressing quick interface
Exit connection11mm (7/16 in.) ID plug-in connection
Output4 0 / 4-20mA output, load impedance 600 ohm
communication method4-20mA output, optional: Modbus RS485, Profibus DP with external controller
size804mm x 452mm x 360mm
Protection levelNEMA 4X/IP56
certificateCE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000)
cETLus (UL 61010-1:2012; NEMA 250:2003; CSA C22.2 No 61010-1:2012)
Australian RCM Marking
weight 
20.5kg (45.2 lb) (no reagents and standards);
30kg (66.2 lb) (including reagents, standards and cleaning agents)
Warranty1 year

 