Welcome Customer !

Membership

Fuzhou Fuguang Water Science&Technology Co.,Ltd
Free Member ·1YRS
FuZhou,FuJian
Manufacturer

Main Products: Ammonia Nitrogen Analyzer|Chlorine Dioxide Analyzer/Residual Chlorine Analyzer|Hardness Tester Ammonia Nitrogen Analyzer|Chlorine Dioxide Analyzer/Residual Chlorine Analyzer|Hardness Tester

Xwboo>Products

Fuzhou Fuguang Water Science&Technology Co.,LtdFree·1YRS

  • E-mail

    fgwater@fuguang.com

  • Phone

    400-668-8627

  • Address

    ADD:Building B, Chenxing Science and Technology Park, No. 168, Jinyan Road, Cangshan District, Fuzhou City, Fujian Province

Contact Now

Phosphaxsc orthophosphate analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
Phosphorus Analyzer
Place of Origin
City, Province
Overview
Phosphaxsc orthophosphate analyzer
Product Details
-Product advantages:
 
Phosphaxsc orthophosphate analyzer uses high-performance colorimetric technology, has the advantages of wide range, high sensitivity, low minimum detection limit, fast response, plug and play, and can continuously determine the concentration of orthophosphate. The function makes the analyzer simple to use and easy to maintain, which greatly reduces the operation and maintenance costs. The measurement results can be displayed in graphical or digital form.
 
-Application industry:
 
The Phosphaxsc orthophosphate analyzer uses high-performance colorimetric technology and can be applied to the detection of orthophosphate in the dephosphorization process of sewage treatment plants.
 
-Instrument features:
 
● Wide measuring range
 
● The minimum detection limit is low, suitable for a variety of sewage treatment applications
 
● Low operation and maintenance costs, reliable analysis methods and measurement results
 
● In controlled nutrient removal process, Phosphaxsc orthophosphate analyzer responds quickly
 
● Easy to install at the measuring point
 
● Low maintenance rate
 
● Equipped with plug-and-play digital controller
 
● Phosphaxsc orthophosphate analyzer can be equipped with Filterprobe sc as sample pretreatment system
Scan the QR code to read on your phone
Technical index
Measurement methodsPhotometric colorimetric method (vanadium molybdenum yellow method)
Measuring range
Range 1: 0.05-15.0 mg / L PO4-P
Range 2: 1.00-50.0 mg / L PO4-P
Minimum detection limit
Range 1: 0.05 mg / L PO4-P
Range 2: 1.00 mg / L PO4-P
Measurement accuracy
Range 1: 2% or 0.05mg / L, whichever is greater
Range 2: 2% or 1.0mg / L, whichever is greater
Reagent consumption
Range 1: every 4 months need to consume 2000mL
Range 2: every 2 months need to consume 2000mL
 
Response time (T90)5min. Including sample preparation
Measurement interval5-120min. Adjustable
FeaturesSelf-cleaning, automatic zero point calibration, comprehensive self-diagnosis. Dual channel models can be selected for continuous sample flow
Ambient temperature-20℃ ~ +45℃
Water temperature+4℃ ~ +40℃
installation
Outdoor-(IP55) and Indoor-Wall-mounted installation, railing installation and vertical installation
 
size(W x H x D)540 × 720 × 390 mm
power supply115V / 50-60Hz or 230V / 50Hz, connected to SC1000 controller
weight35kg, including reagents

 