Filtrax sampling pretreatment system

Model
Nature of the Manufacturer
Manufacturer
Product Category
Sampling Preprocessing
Place of Origin
City, Province
Overview
Filtrax sampling pretreatment system
Product Details
typical application
Suitable for sample pretreatment of online instruments.
 
Features and benefits
● Using ultrafiltration technology, ultrafiltration membrane can filter 0.15μm particles;
● Two peristaltic pumps work alternately, taking samples in turn;
● Unique air cleaning design, can automatically clean its built-in filter membrane, to minimize cleaning work;
● Submersible pumps that do not require expensive and often require maintenance
● The sampling tube can be heated to ensure that Filtrax can be used outdoors in any weather conditions.
● The system can automatically monitor the sample flow rate
 
Principle of operation
Filter membrane A and filter membrane B made of special polymer materials are placed in the same stainless steel container and immersed directly in the sampling water. The filter membrane A and the filter membrane B are drawn into the transfer pipe by their respective samples, and connected to the peristaltic pump A and the peristaltic pump B in the controller. The two peristaltic pumps work alternately; during the operation of a peristaltic pump, the sample is filtered by the corresponding filter membrane, drawn into the controller, and then transmitted to the subsequent online water quality analyzer. During the entire working process of the Filtrax sample pretreatment system, the air compressor inside the controller continuously works, and the generated compressed air passes through two air transmission pipes and is transmitted to the exhaust hole at the bottom of each filter membrane; When one of the peristaltic pumps stops working, the suspended particles adsorbed on the surface of the corresponding filter membrane are removed from the surface of the filter membrane. So as to ensure that Filtrax sample pretreatment system can work continuously.
Scan the QR code to read on your phone

 

Technical index
Sample flow rateAbout 900mL / h
Sample flow rate230VAC±10%, 50-60Hz
Sample temperature5~40℃
Ambient temperature-20~40℃
Chassis levelIP 55 (outdoor installation)
Instrument size
Control unit: 430 × 530 × 220mm
Filter container: 92 × 500 × 340mm
weight41kg
Sample suction tube length5 meters (heated)
Optional sample transfer tube
2 meters (without heating); 10 meters (with heating);
20 meters (heating); 30 meters (heating)
 

 