The Walk-in test chamber simulates the effects that a range of temperature and humidity conditions have on a product or material, for example, telecommunications equipment or photovoltaic solar panels that are exposed to outdoor temperature and humidity conditions.
Applications
KOMEG Walk in test Chamber are used for various types of testing such as prototype evaluation, research & development testing, production testing, accelerated stress testing, reliability testing, failure analysis, and any other application where environmental simulation is needed.
Features
Walk-in, Drive-in Type
Stainless steel interior (or galvanized for savings)
Painted galvanized steel exterior for long life
MAP service doors are hinged (no tools or lifting)
Easy access to components
Reliable steam generator system for humidity applications
Sheathed electric heaters for safety and reliability
Adjustable vanes in supply grille allow directing of airflow as needed
Standards Implemented and met
GB-2423. 1-89(IEC68-2-1) Test A: Low Temperature Test
GB-2423. 2-89(IEC68-2-2) Test B: High Temperature Test
GJB360. 8-87(MIL-STD.202F) High Temperature Operating Life Test
GBJl50. 3(MIL-STD-810D) High Temperature Test
GJBl50. 4(MIL-STD-810D) Low Temperature Test
GB2423. 3-93(IEC68-2-3) Test Ca: method for steady-state damp heat test
GB2423. 4-93(IEC68-2—30) Test Db: Damp Heat Alternative Test
GJBl50. 9-93(MIL-STD-810D) Damp Heat Test
| Main Technical Index |
| Air-cooled. The specifications measurement in the environment temperature at + 25 ℃ without load |
| 1. Temperature |
| 1.1 Temp. range | -40℃~+150℃ |
| 1.2 Temp. deviation | ≤±2.0℃ |
| 1.3 Temp. fluctuation | ±0.5℃ |
| 1.4 Heating up rate | From -40 ℃ to + 150 ℃ 2℃/min non-linear no load |
| 1.5 Cooling rate | From +150 ℃ to -40 ℃ 2℃/min non- linear no load |
| 2. Humidity |
| 2.1 Humidity range | 25%R.H.~98%R.H |
| 2.2 Humidity range | |
| 2.3 Humidity deviation | ±3.0%RH(>75%RH)±5.0%RH(≤75%RH) |
| 2.4 Humidity fluctuation | ±2.0%RH |
| 4. Other requirements | Humidity range adjustment possible between +25 to +55 degreesDew point: +25 to 54 Celsius |
Temperature and humidity performance test is in accordance with the relevant provisions of the IEC60068 -3 standard measurement; sensor placed in the unit outlet.According to national test standard: GB/T 5170.2, use 9 test point calibration, as below: |
| Body Structure |
Overall structure and chamber is composed of three parts as below.Insulation box, separate refrigeration units, and electrical control cabinet. |
| 1. Body dimensions | Workplace volume: W 2500 ×H 2000 ×D 2500 mm Exterior size: W 4350 ×H 2400 ×D 2800 mm |
| 2. Insulation Box | ※ wall material: high-quality carbon steel※ inner wall material: SUS304 # matte stainless steel plate※ Insulation materials: 100mm rigid polyurethane foam insulation layer + glass fiber. |
| 3. Door | Double door W(750+750)×H 2000mm, with observation window (the size depend on final design) |
| 4. Lighting Device | Moisture-proof lighting, with explosion-proof lampshades. |
| 5. Cable port | φ100mm*1 located on both side of chamber, with rubber stopper |
| 6.Electric control box | Total power circuit breaker, over-temperature protection. |
| 7. Control model | Forced circulation ventilation, balance thermostat (BTHC) |
| 8. Heating | High quality nickel-chromium alloy wire electric heaters,Non-contact control mode(SSR) |
| 9. Water outlet hole | Available for drain the condensate water |
| 10. Water purifier | Pure water device, 1 pcs, please provide 1 ~ 2 kg/cm2 tap water |
| 11. Water supply system | Pump to enhance automatic water supply |
| 12. Humidification | Stainless steel corrosion-resistant electric evaporative humidification |
| 13. Humidity sensor | High accuracy Vaisala electronic humidity sensor |