2 - ZONE THERMAL SHOCK TEST CHAMBER IS EQUIPPED WITH A HOT AND A COLD DUAL ZONE WITH INDIVIDUAL CHAMBER CONTROL. IT PROVIDES A THERMAL SHOCK RANGE FROM (-80 ℃ TO 200 ℃).
Thermal shock testing also called temperature shock testing or temperature cycling exposes products toalternating low and high air temperatures to accelerate failures caused by repeated temperature variations during normal use conditions. It is used to accelerate failure modes caused by very rapid changes in temperature.
Two-zone thermal shock test chamber is equipped with a hot and a cold dual zone with individual chamber control. It provides a thermal shock range from (-80 ℃ to 200 ℃).
| 2 Zone Thermal Shock Test Chamber |
| ALL THE MODEL OF THIS KIND PRODUCTS IN THE BLANKET: |
| Specifications |
| Model | TST-A Series | TST-B Series | TST-D Series |
| 64A | 64B | 64D |
| Test weight(KG) | 410 | 410 | 410 |
| Interior size | W(mm) | 400 | 400 | 400 |
| H(mm) | 400 | 400 | 400 |
| D(mm) | 400 | 400 | 400 |
| Basket size | W(mm) | 860 | 860 | 860 |
| H(mm) | 2070 | 2070 | 2070 |
| D(mm) | 2000 | 2000 | 2000 |
| Low Temp. chamber | -55 °C~ -10°C | -70 °C~ -10°C | -80 °C~ -10 °C |
| Pre-cooling time | +20 °C~ -55°C,Within 60 min | +20°C~ -70 °C,Within 60 min | +20 °C~ -80°C,Within 60 mins |
| Test. chamber | -40 °C~ +150°C | -55°C~ +150°C | -65°C~ +150 °C |
| High Temp. chamber | +60 °C~ +200 °C |
| Pre-heating time | +60°C~ +200°C ,Within 25 min |
| Recovery time | High-Temp Exposure 30 min |
| Low-Temp Exposure 30 min |
| Recovery time within 5 min |
| Safety device | within 10 seconds |
| Standard accessory | No fuse breaker, over pressure over heat and over current protection for compressor, over temperature protection, over load protection for blower, dry heat protection |
| Interior material | Cable port(portΦ50mm)*1, shelves*2 |
| Exterior material | Stainless steel plate ( SUS 304 ) |
| Insulation material | Baked painting steel or stainless steel(SUS304) |
| Refrigeration system | Rigid polyurethane Foam glass fiber wool |
| Air Cooled |
| Controller | Panel:7-inch LCD Touch panel,Chinese or English display selectable |
| Operation model:Program or fix point running |
| Memory capacity:120programs,1200steps,all repeat 999cycles,part repeat 99 cycles. |
| Accuracy:Temp.:0.1% of F.S±1 digit Humidity:0.1% of F.S±1 digit |
| Input:Temp.:pt-100 Humidity:pt-100 or ~5V DCV |
| Ambient Temp. | +5ºC~+35ºC |
| Power | AC 380±10%50Hz ,3 phase 4 wires +Ground Wires |
| Note:1. We reserve the right to change specifications without prior notice 2. Customized sizes and configurations available |