Rapid temperature change chamber are very powerful climatic chambers used to perform hot and cold cycles with steep temperature gradients(from 3 to 15C/min depending on the temperature ranges)
Rapid Temperature Change Chamber / 3000 L
Standards Implemented and met
GB-2423.1-2008(IEC68-2-1)Test A: Cold Test
GB-2423.2-2008(IEC68-2-2)Test B: Dry Heat Test
GJB360.8-2009(MIL-STD.202F) Heat Aging Test
GJBl50.3-2009 (MIL-STD-810D) high temperature test method.
GJBl50.4-2009 (MIL-STD-810D) low temperature test method.
GB2423.3-2008 (IEC68-2-3) Test Ca: Constant damp heat test method.
GB2423.4-2008 (IEC68-2-30) Test Db: Alternating damp heat test method.
GJBl50.9-2009(MIL-STD-810D) Damp heat test chamber technical conditions
| Main Technical |
| Water-cooled. The specifications measurement in the environment temperature at + 25 ℃ without load |
| volume | 3000 L |
| 1. Temperature |
| 1.1 Temp. range | -70℃~+150℃ |
| 1.2 Temp. deviation | ≤±2.0℃ |
| 1.3 Temp. fluctuation | ±0.5℃ |
| 1.4 Heating up rate | From -40℃ to +85℃, linear, 15℃/min |
| 1.5 Cooling rate | From -40℃ to +85℃, linear, 15℃/min |
| 2. Humidity |
| 2.1 Humidity range | 20%R.H.~98%R.H |
| 2.2 Humidity range | |
| 2.3 Humidity deviation | ±3.0%RH(>75%RH)±5.0%RH(≤75%RH) |
| 2.4 Humidity fluctuation | ±2.0%RH |
| 3. Floor capacity | 300KG/m2 |
| 4. Full load | 10 PV Modules to be loaded in a vertical orientation. Single panel max. size of 2000*1000*50mm. Total weight of 300KG. |
Temperature and humidity performance test is in accordance with the relevant provisions of the IEC60068 -3 standard measurement; sensor placed in the unit outlet.According to national test standard: GB/T 5170.2, use 9 test point calibration, as below: |
| Body Structure |
Overall structure and chamber is composed of three parts as below.Insulation box, separate refrigeration units, and electrical control cabinet. |
| 1. Body dimensions | Workplace volume: W 1000 × H 1500 × D 2000 mm,Exterior size: W 1500 × H 3500 × D 4350 mm, |
| 2. Insulation Box | ※ wall material: high-quality carbon steel※ inner wall material: SUS304 # matte stainless steel plate※ Insulation materials: 100mm rigid polyurethane foam insulation layer + glass fiber. |
| 3. Door | Double door (according to design), with observation window (the size depend on final design) |
| 4. Lighting Device | Moisture-proof lighting, with explosion-proof lampshades. |
| 5. Cable port | φ100mm*1 located on both side of chamber, with rubber stopper |
| 6.Electric control box | Total power circuit breaker, over-temperature protection. |
| 7. Control model | This device is Cascade compression refrigeration We use intellectual cooling control: PID will control solenoid-valve and heater on/off to adjust cooling or heating capacity according to temperature and loading requirements and status inside the chamber. Only cooling or only heating will be activated to get temperature balance to save energy |
| 8. Heating | High quality nickel-chromium alloy wire electric heaters,Non-contact control mode(SSR) |
| 9. Water outlet hole | Available for drain the condensate water |
| 10. Water purifier | Pure water device, 1 pcs, please provide 1 ~ 2 kg/cm2 tap water |
| 11. Water supply system | Pump to enhance automatic water supply |
| 12. Humidification | Stainless steel corrosion-resistant electric evaporative humidification |
| 13. Humidity sensor | High accuracy Vaisala electronic humidity sensor |