Welcome Customer !

Membership

HANWEI ELECTRONICS GROUP CORPORATION
Free Member ·1YRS
ZhengZhou,HeNan
Manufacturer

Main Products: Gas Detector

Xwboo>Products

SC-odor600 Fixed Source Volatile Organic Gas Online Monitoring System

Model
Nature of the Manufacturer
Manufacturer
Product Category
Environmental Parameter Monitoring Devices
Place of Origin
City, Province
Overview
Analytical method: Gas Chromatography, GC+Dual FIDsampling method: Direct extraction + full heat methodFID Baseline Noise: Baseline noise ≤ 5×10 -14 AFID baseline drift: ≤ 3×10 -13 A / 30minFID detection limit: ≤ 2.9×10 -12 g/s
Product Details

Description

SC-odor600 fixed source volatile organic gas online monitoring system consists of flue gas sampler, temperature pressure flow humidity detector, heating cable, analysis cabinet pre-treatment system, auxiliary gas source, gas analyzer, data acquisition and processing, data transmission, etc. SC-odor600 fixed source volatile organic gas online monitoring system, which can continuously monitor the concentration of various organic gases such as total hydrocarbons, methane, non-methane total hydrocarbons, benzene, toluene, xylene, etc. or other gas concentrations of VOCs components. As well as gas temperature, pressure, flow rate, humidity and other related parameters, as well as statistical emission rate, total emission and other parameters, and can effectively manage the acquired data.

Features

Fully consider the sample gas conditions, dew point and other factors, and design the pretreatment system in a targeted manner

Explosion -proof/non-explosion-proof type optional

Highly anti-corrosion design. All parts that come into contact with the sample (for example, piping, valves, pressure regulators, and flow meters)

Special leak-proof design of sealing components

Based on the design concept of anti-corrosion and anti-leakage to ensure long-term reliable continuous operation of the system

Anti -blocking design of sampling system based on automatic high-frequency and high-pressure back flushing method, reducing maintenance

Advanced system technology, stable and reliable, strong applicability

On the basis of meeting environmental protection requirements, it is economical and reasonable

The system design fully considers the characteristics of environmental protection, and can expand the measurement requirements to improve the utilization value of the system

The analysis method is consistent with the standard method, and the GC-FID method is used to detect

Maintenance -free design for continuous operation, automatic cycle operation after startup, true online analysis

Simultaneous online detection of components such as non-methane total hydrocarbons and benzene series

Application

Petrochemical industry
Fine chemical industry
Biopharmaceutical industry
Rubber products industry
Printing and packaging printing industry
Textile printing and dyeing industry
Spray coating workshop
Electronic semiconductor industry
Synthetic /artificial leather industry
Efficiency monitoring of waste gas treatment device

Specifications

Analytical methodGas Chromatography, GC+Dual FID
Sampling methodDirect extraction + full heat method
FID Baseline NoiseBaseline noise ≤ 5×10 -14 A
FID baseline drift≤ 3×10 -13 A / 30min
FID detection limit≤ 2.9×10 -12 g/s
FID detection limitNon-methane total hydrocarbons: 0.05mg/m3
Benzene series: 0.05mg/m3
RangeNon-methane total hydrocarbons: 0-200mg/m3
Benzene series: 0-40mg/m3
Indication error≤ ±2% FS
Repeatability bias≤ 3%
Zero drift≤±1% FS
Range drift≤±2% FS
Signal output4-20mA, RS232, RS485
Sample pump flow5L/min
Gas consumptionH2 (30±10)ml/min
Air consumptionAir (400±20)ml/min
Carrier gasNitrogen, more than 99.999% pure
Carrier gas consumptionN2 (80±10)ml/min
Insulation resistanceNot less than 20MΩ
Similar Product Recommend