CX-ICM-1104 industrial conductivity meter

Model
Nature of the Manufacturer
Other
Product Category
On-line Instrument
Place of Origin
City, Province
Overview
Illustrate:It can be widely used for continuous monitoring of the conductivity values of the solutions used in the following industries: thermal power plants, chemical industry, chemical fertilizer, metallurgy, environmental protection, pharmacy manufacturing, biochemistry, foodstuff and running water.
Product Details

CX-ICM-1104 industrial conductivity meter

Main features
1.Fully intelligent: It adopts high-precision AD conversion and single chip microcomputer processing technologies
2.It can be used for conductivity and temperature measurement, automatic temperature compensation, automatic range switching, self-checking, etc.
3.Multi-parameter display
4.values with red 10 x 10mm Nixie tubes. As red display is eye-catching, the displayed values can be seen from a long distance away.
5.Strong interference immunity: Its current output adopts optoelectronic isolating technology. This meter has strong interference immunity and the capacity of long-distance transmission as well as excellent electromagnetic compatibility.
6.25C Conversion: The conductivity value measured at the current temperature is converted to that at 25C; thus, this conductivity meter always displays the conductivity value at 25C. Therefore, this meter is especially suitable for testing the various water solutions in power plants.
7.Automatic range switching: The measuring ranges can automatically switch within the measuring range of the electrode.
8.Communication interface: It can be easily connected to a computer to perform monitoring and communication.
9. With strong environmental adaptability, clear display, easy operation and excellent measuring performance

Main parameters

StandardJB/T6855-93 “Industrial Conductivity Meter”
Conductivity measuring range0.0~20µS/cm(Electrode: K=0.01),0.1~200µS/ cm (Electrode: K=0.1)
1.0~2000µS/cm(Electrode:K=1.0),10~20000µS/cm(Electrode: K=10.0)
AccuracyConductivity±1.0%FS; temperature: ±0.5C
Automatic temperature compensation error±0.5%FS
Intrinsic error of the instrumentconductivity: ±1.0%F.S, temperature: ±0.5C(0~60C)
Intrinsic error of electronic unitsconductivity: ±0.5%F.S, temperature: ±0 3C(0~60C)
Range for automatic temperature compensation0~80C, with 25C as the reference temperature
Water sample tested0~80C, 0.6Mpa
Repeatability error±0.2%FS±1 unit
Stability±0.2%±l unit FS/24H
Alarm error≤±3.0%FS
Error caused by ambient temperature≤±0.5%FS
Error by the supply voltage≤±0.3%FS
Isolated current output4~20mA (load<750kΩ)
Output current error≤±l.0%FS
Alarm relayAC220V, 3A
Communication interfaceRS485 or 232(optional)
Power supplyAC 220V±22V, 50Hz±1Hz
ProtectionIP65
Overall dimension146 (length) x 146 (width) x 108 (depth) mm
dimension of the hole138 x 138mm
Working conditionsAmbient temperature: 0~60C; relative humidity <85%
Connection tubes for inlet and outlet waterΦ8, Φ10 and Φl2
Note:This meter can be equipped with the following four kinds of conductivity electrodes: K=0.01, K=0.1, K=1.0 ,or K=10.0