Welcome Customer !

Membership

Cixi flowmeter Co., Ltd.
Free Member ·1YRS
,
Other

Main Products: 流量计

Xwboo>Products

CX-LSM-1604 lab sodium meter

Model
Nature of the Manufacturer
Other
Product Category
Lab Instrument
Place of Origin
City, Province
Overview
Illustrate:It can be widely used for the continuous monitoring of sodium ions in the solutions of the thermal power plants, chemical fertilizer, metallurgy, environmental protection, pharmaceutical, biochemical, foodstuff and running water.
Product Details

CX-LSM-1604 Lab Sodium Meter


Brief Introduction
1.CX-LSM-1604 is a sodium meter with the most powerful functions and the most convenient operation on the market.
2.It is equipped with the dedicated laboratory ppb sodium ion measurement electrode and reference electrode.
3.The performance of the static measurement in the beaker has been greatly improved
 
Main Features
1.Computer-based It adopts CPU high-performance chip, high-precision AD conversion technology and SMT technology to complete multi-parameter measurement, temperature compensation and instrument self-checking
2.it is also featured by high precision and good repeatability.
3.High reliability Single board structure, touch button, no switch knob; input impedance> 1012Ω.
4.High Sensitivity In the static water sample, it can measure 0.2ppb (8pNa); With a long stablization time, it has very stable readings.
5.It is of waterproof and anti-erosion design.
6.Electrode diagnosis With a variety of calibration modes and EO and S queries, it conducts the monitoring of electrode performance.
7.Network function RS232 bi-directional communication, which can be equipped with a printer or TLab8.0communications and data processing software.
8.Historical data 3000 data can be stored.
9.Notepad 200 messages can be recorded.
10.Built-in real-time clock Providing a variety of functional time references.
 
Main Parameters

Measurement range0.2 ug/L~23000mg/L; 0~8.00 pNa, -1000~ +1000 mV 0~99.9°C
Resolution 0.1 ug/L; 0.1 mV,0.1°C
Repeatability Reading ± 2.5%
Accuracy±0 2 mV, ±0 3°C
Continuity stability time>100 seconds
Detection lower limit0.2 ug/L (8pNa); Input impedance >1012Ω
Temperature compensation rangeautomatic/manual 0-99.9 °C
Response speed32 seconds (25 °C) or 50 seconds (1 5°C)
Communication interfaceRS232 bi-directional communication
Power supplyAC220V±22V, 50Hz ±1Hz
Working conditionsAmbient temperature 0-60°C; relative humidity ≤90
power-down memory> 5 years
Clock accuracy± l minute/month
Calibration modeOne-point calibration, two-point calibration, manual input of EO and s
Note: Standard configuration ;One secondary instrument; one power supply; one reference electrode, one measuring electrode, one temperatureelectrode; a set of electrode support, a set of standard buffer solution