Waveform of tested signal: Sine wave、β=0.02Frequency of tested signal: (50±0.2)HzPosition of tested lead in jaw: no special requirementsAmplitude range of tested signal when test phase: 100~220V、0.5A~1.5AExternal frequency electromagnetic interference: should avoidWaveform of tested signal: sine、β=0.05Frequency of tested object: (50±0.5)HzPosition of tested lead in jaw: no special requirementsAmplitude range of tested signal when test phaseWhile measure U1-U2: 30V~500VWhile measure I1-I2 : 10mA~10.00AWhile measure ......
Product Details
Brief introduction
This digital phase meter is designed for field measurement of voltage, current and phase. It’s handheld, with high precision, low cost, and double channel input.
This Digital Phase Meter can be used to measure the phase between U-U, I-I and U-I in the field, distinguish the inductive and capacitive circuit and phase sequence of three-phase voltage, detect the wiring group of transformer, test the secondary circuit and the mother differential protection system, and read out the phase relationship between the CT of differential protection groups,check the wiring of the watt-hour meter, etc
The clamp current transformer is used to input the measured current, so there is no need to disconnect the measured line during the measurement.
When measuring the phase between U1-U2, the two input circuits are completely insulated and isolated, so that the short-circuit of the tested line caused by possible miswiring can be completely avoided, to avoid meter burning .
High contrast LCD screen, words up to 25mm, screen Angle can be freely converted to about 70°, to obtain the best visual effect.
The Digital Phase Meter shell is made of engineering insulating material and rubber anti-vibration protective cover, which is safe and reliable.
Basic error
1.Reference working environment
1)Temperature: (23±5)℃
2)Humidity: (45~75)% RH
3)Waveform of tested signal: Sine wave、β=0.02
4)Frequency of tested signal: (50±0.2)Hz
5)Position of tested lead in jaw: no special requirements
6)Amplitude range of tested signal when test phase: 100~220V、0.5A~1.5A
7)External frequency electromagnetic interference: should avoid
2.Basic error limit
1)AC voltage
Voltage
Resolution
Error limit
20V
0.01V
±(1.2%RD+2)
200V
0.1V
±(1.0%RD+2)
500V
1V
±(1.2%RD+2)
Input impedance:2MΩ for each one
2)AC current
Current
Resolution
Error limit
200mA
0.1mA
±(1.0%RD+2)
2A
1mA
10A
10mA
3) Phase U-U、U-I、I-I
Range
Resolution
Error limit
0~360°
1°
±3°
The voltage input loop impedance when measured in U1-U2 phase:40kΩ
Rated working environment
1)Environment temperature: (0~40)℃
2)Humidity: (20~80)% RH
3)Waveform of tested signal: sine、β=0.05
4)Frequency of tested object: (50±0.5)Hz
5)Position of tested lead in jaw: no special requirements
6)Amplitude range of tested signal when test phase
While measure U1-U2: 30V~500V
While measure I1-I2 : 10mA~10.00A
While measure U1-I2 or I1-U2: 10V~500V、10mA~10.00A
7)External frequency electromagnetic interference: should avoid
Rated working error limit of Digital Phase Meter
Under above rated working environment, the measured rated working error should not beyond two times related basic error.
Other technical specification
1.Digits: three digits and a half
2.Sample rate: 3 times/second
3.Power: single 9V battery, current less than 5mA
4.Size
Shell: 192mm×95mm×55mm
Clamp shell: 140mm×42mm×20mm
Jaw: Φ7mm×9mm
More pictures about double clamp voltmeter
F.A.Q
1.Are you manufacturer or trading company ?
We are ISO certified manufacturer .
2.How about the warranty ?
One year free and lifelong maintenance .
3.Can you afford calibration certificate ?
Yes , we can afford calibration certificate issued by third authoritative institution .
4.Do you have Technical Backup to all customers from different countries ?
Sure, we have rich-experienced engineer service you within 24 hours through phone call , ,or on-line communication to solve your problems . And if if necessary , our engineer can visit your company to afford technical support face to face .
Company profile
Wuhan Goldhome Hipot Electrical Co.,ltd was founded in 2008 ,as predecessor is the State owned tertiary Enterprises enterprises. Main products included Primary business was sale, design and manufacture high-voltage transmission and insulation test set. hipot tester, GIS PD test system, AC/DC high-voltage test transformer, primary current injection test set,insulation resistance test meter, Insulation withstand voltage test device and other testing instruments.