Phase current range applicable:1A±0.1A(Reference Conditions)10mA~20.0A(Operating Conditions)Phase voltage range applicable:100V±10V(Reference Conditions) 30V~600V(Operating Conditions) Function:Directly test phase, A.C. current, A.C. voltage, A.C. leak current, and phase sequence; identify grouping, inductive, and capacitive circuit of transformer; and indirectly test power factor and power.
Product Details
Introduction
HM 2000B dual-clamp Digital Phase Tester is a multi-functional instrument developed specially for on-site test. Featured with high precision, good stability, less power consumption and easy operation, it measures directly A.C. voltage and A.C. current even when the tested circuit is not opened, as well as phase between two voltages, two currents and the voltage-current. In addition, it measures indirectly power factor and power of the circuit, identifies three-phase sequence, wiring group, inductive and capacitive circuit of transformer, tests secondary circuit and the busbar differential protection system, tells the phase relation between different CT groups of differential protection, examines whether the wiring of watt-hour meter is correct, and checks and repairs circuit equipments, etc. The instrument has huge LCD screen, with letters reaching 40mm in height and looks greatly clear thanks to the backlight design. The instrument is applicable to electricity, petrochemical engineering, metallurgy, railway, meteorology, industrial and mining enterprises, academic institutions, colleges and metrological sectors, etc.
Main Parameter
2.1. Reference Conditions and Operating Conditions
Influence Factors
Reference Conditions
Operating Conditions
Remarks
Ambient environment
23℃±1℃
-10℃~40℃
Ambient humidity
40%~60%
<80%
Signal waveform
Sine wave
Sine wave
β=0.05
Signal frequency
50HZ±1HZ
45HZ~65HZ
Working voltage
9V±0.1V
9V±1V
Phase current range applicable
1A±0.1A
10mA~20.0A
Phase voltage range applicable
100V±10V
30V~600V
External electric field and magnetic field
Shall be avoided
Location of tested wire
The tested wire is located in approximate geometric center of the clamp jaw
2.2. General Specifications
Function
Directly test phase, A.C. current, A.C. voltage, A.C. leak current, and phase sequence; identify grouping, inductive, and capacitive circuit of transformer; and indirectly test power factor and power.
Power Supply
DC9V alkaline dry batteries (1.5V AA×6)
Power Consumption
Nearly 35mA at the maximum if backlight on, batteries work continuously for nearly 40 hours.
Nearly 5mA if backlight off, batteries work continuously for 300 hours.
Display
LCD display, blue screen with backlight design, applicable to dark places.
LCD Size
70mm×62mm
LCD Display Field
64mm×54mm
Instrument Size
(L x W x H): 196mm×92mm×54mm
Jaw Size:
φ7.5mm×13mm
Sampling Rate
Nearly 3 times/second
Range
A.C Voltage: 0~20V/200V/600V
A.C Current: 0~200mA/2A/20A
Phase: 0~360°
Range of tested signal during phase measuring
When measuring phase U1-U2: 30V~600V
When measuring phase I1-I2: 10mA~20.00A
When measuring U1-I2 or I1-U2: 10V~600V, 10mA~20.00A
Auto power-off
The instrument shut off automatically in 15min after startup so that to reduce battery consumption.
Voltage test
When the battery voltage is less than 7.2v, the low voltage symbol appears, reminding to replace the battery.
Mass of instrument
The body is nearly 550g (battery included), the clamp is nearly 170g x2, and the test wire is nearly 250g.
Length of test wire
1.5m
Line length of current clamp
2m
Operating temperature and humidity
-10℃~40℃; 80%Rh and less
Storage temperature and humidity
-10℃~60℃; 70%Rh and less
Input impedance
Input impedance of tested voltages: 2MΩ
Input impedance of voltage when testing phases U1U2: 40kΩ
Withstand voltage
Withstand voltage between upper cover of the instrument and bolt on battery cover place is 2000V/50Hz sine wave A.C. current for one minute.
Insulation
Insulation between wires and the shell of instrument and insulation between two voltage input ends ≥10MΩ.
Structure
Double insulation
2.3. Basic Errors and Performance Indexes under Reference Conditions
Category
Range
Resolution
Basic Error
Voltage
20V
0.01V
±(1.2%rdg+2dgt)
200V
0.1V
500V
1V
Current
200mA
0.1mA
±(1.0%rdg+2dgt)
2A
1mA
10A
10mA
Phase
0~360°
1°
±1°
Note: the phase error under operating conditions shall be no greater than ±3°.
Structure
1. current input interface
2. Product model label place
3. LCD
4.“ON" button for startup
5.“HOLD" button for data-hold
6. Knob switch for functions
7.Voltage input jack
8. Clamp jaw
9. Current clamp
10.Trigger
11. Current clamp lead
12. current input interface
13.“OFF" button for shut-off
14. “LAMP" button for backlight
15.Voltage test clip
16.Voltage test wire
Reference drawings for test wiring
1: Test single-phase U-I phase
Identify inductive or capacitive circuit
2: Test three-phase four-wire Ua-Ib phase
3: Test three-phase four-wire Ia-Ib phase
4: Test three-phase four-wire Ua-Ib phase
Identify the three-phase four-wire phase sequence
5: Test three-phase three-wire Uab-Ucb phase
Identify the three-phase three-wire phase sequence