Sf6 Gas Analyzer

Model
Nature of the Manufacturer
Manufacturer
Product Category
Sf6 Analyzer
Place of Origin
City, Province
Overview
HM3010 SF6 gas analyzer is a high-precision intelligent instrument for measuring trace moisture content in SF6 gas. The self-locking joint of the instrument adopts the self-locking joint imported from Germany, which is safe and reliable without air leakage.
Product Details
Product overview


HM3010 SF6 gas analyzer is a high-precision intelligent instrument for measuring trace moisture content in SF6 gas. The self-locking joint of the instrument adopts the self-locking joint imported from Germany, which is safe and reliable without air leakage.


Technical feature


1.Portable design: The instrument is lighter, easy to carry and use.

2.Fast measurement: No need to wait after the instrument is turned on, measure immediately and get the humidity value quickly.

3.Fast gas saving: Gas consumption is only about 2L (101.2kPa) during measurement.

4.Self-locking joints: The self-locking joints imported from Germany are used, which are safe and reliable without air leakage.

5.Data storage: adopts large-capacity design and can store up to 50 sets of test data.

6. Clear display: The LCD screen directly displays the dew point, micro water (ppm), ambient temperature, ambient humidity, time and date, battery power, etc.

7.RS232 interface: It can be connected to a serial printer for data printing.

8.Built-in power supply: Built-in large-capacity rechargeable lithium battery, which can work continuously for 10 hours after a full charge.


Parameter table


Dew point

Measurement range

-60 ℃~+20 ℃

Measurement Accuracy

±2℃(-60℃~+20℃)

Measurement time

<3min.

Temperature

-40℃~+60℃

Humidity

0~100% RH

Power

AC 220V

Built-in rechargeable battery

Weight

3kg

Size

250×100×300mm3

Operating temperature

-30℃~+50℃


1


Measurement methods


1. Connect SF6 equipment

Connect the threaded end of the measuring pipe to the switch connector, tighten it with a wrench, and close the needle valve on the other end of the measuring pipe;

Then insert one end of the quick connector on the test pipe into the sampling port on the micro water meter;

Connect the exhaust pipe to the air outlet.

Finally, connect the switch connector to the SF6 electrical equipment measurement interface, and tighten it with a wrench;

2. Check the battery

It is recommended to use AC power first.

When using DC power, please check the battery power displayed in the upper right corner. If the power is less than about 20%, please turn off the battery and continue to use it after charging.

3. Start measuring

First open all the needle valves on the measuring pipe of the instrument, and then use the flow valve on the panel to adjust the flow, adjust the flow to about 0.5L/M, and start measuring the SF6 dew point.

The measurement time of the first device takes 5-10 minutes, and each device thereafter takes 3-5 minutes.

4. Store data

After the equipment measurement is completed, you can save the data in the instrument and press the "OK" key to call up the operation menu. For specific operation methods, see the next section.

5. Measure other equipment

After a piece of equipment measures, close the needle valve on the measuring pipe and the regulating valve on the micro water meter. Remove the adapter from the SF6 electrical equipment. If you need to continue to measure other devices, follow the steps above to continue to measure the next device.

6. End of measurement

After the measurement of all equipment is finished, turn off the power of the micro water meter.


Precautions


If you are using this product for the first time, please note the following:

1.To start the measurement, first open all the needle valves on the measuring pipe of the instrument, and then adjust the flow with the flow valve on the panel. When the measurement is over, the operation is reversed.

2.When the micro-water meter is not used for a long time, there will be some air left in the test pipe and air chamber. Therefore, in the first test, the high humidity gas in it will affect the test speed, because the tested SF6 gas needs to reach saturation. First take away the moisture in the air. So we will find that when testing the second SF6 electrical equipment and the measurement speed will be very fast (3-5 minutes).