Welcome Customer !

Membership

Biobase Biodusty(Shandong), Co., Ltd.
Free Member ·1YRS
,
Other

Main Products: 实验室设备

Xwboo>Products
Product Categories

Biobase Biodusty(Shandong), Co., Ltd.Free·1YRS

  • E-mail

    export@biobase.com

  • Phone

    +86-531-81219803

  • Address

    9 Gangxing road, High-tech Zone, Jinan, Shandong, China

Contact Now

MicroPette Plus Autoclavable Pipette

Model
Nature of the Manufacturer
Other
Product Category
Place of Origin
City, Province
Overview
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.
Product Details

All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22℃.



Fully Autoclave:

The MicroPette plus pipettes can be fully autoclaved at 121℃, 1 atm for minutes. The pipettes can be autoclaved without special preparations. After autoclaving the pipette must be cooled down and dried over 12 hours.


Features for MicroPette plus:

* Fully autoclavable

* Light weight, ergonomic, low force design 

* Digital display clearly reads volume setting

* The pipettes cover volume range of 0.1μl to 10ml

* Easy to calibrate and maintain 

* Design helps avoid repetitive strain injuries

* Manufactured from innovative materials




Features for Multi-channel pipette

* 8 and 12 channel pipettes are available for standard 96-well plate 

* Individual piston and tip cone assemblies allow easy repair and maintenance 

* Compound material tip cone design allows visual seal vertification

* Universal pipette tips can be used 

* Dispensing head rotates ensure optimum pipetting convenience 


Models:


Single-channel Adjustable Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

   (Inaccuracy)

Maximume Permissible

Random error

   (Inaccuracy)




%

μl

%

μl

0.1-2.5μl

0.05μl

2.5

2.50%

0.0625

2.00%

0.05

1.25

3.00%

0.0375

3.00%

0.0375

0.25

12.00%

0.03

6.00%

0.015

0.5-10μl

0.1μl

10

1.00%

0.1

0.80%

0.08

5

1.50%

0.075

1.50%

0.075

1

2.50%

0.025

1.50%

0.015

2-20μl

0.5μl

20

0.90%

0.18

0.40%

0.08

10

1.20%

0.12

1.00%

0.1

2

3.00%

0.06

2.00%

0.04

5-50μl

0.5μl

50

0.60%

0.3

0.30%

0.15

25

0.90%

0.225

0.60%

0.15

5

2.00%

0.1

2.00%

0.1

10-100μl

1μl

100

0.80%

0.8

0.15%

0.15

50

1.00%

0.5

0.40%

0.2

10

3.00%

0.3

1.50%

0.15

20-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

20

3.00%

0.6

1.00%

0.2

50-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

50

1.00%

0.5

0.40%

0.2

100-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

100

2.00%

2

0.70%

0.7

200-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

200

0.90%

1.8

0.30%

0.6

1000-5000μl

50μl

5000

0.50%

25

0.15%

7.5

2500

0.60%

15

0.30%

7.5

1000

0.70%

7

0.30%

3




8-channel Adjustable Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

  (Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)




%

μl

%

μl

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15

5

2.50%

0.125

2.50%

0.125

1

4.00%

0.04

4.00%

0.04

5-50μl

0.5μl

50

1.00%

0.5

0.50%

0.25

25

1.50%

0.375

1.00%

0.25

5

3.00%

0.15

2.00%

0.1

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75

150

1.00%

1.5

0.50%

0.75

50

1.50%

0.75

0.80%

0.4

12-channel Adjustable Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

   (Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)




%

μl

%

μl

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15

5

2.50%

0.125

2.50%

0.125

1

4.00%

0.04

4.00%

0.04

5-50μl

0.5μl

50

1.00%

0.5

0.50%

0.25

25

1.50%

0.375

1.00%

0.25

5

3.00%

0.15

2.00%

0.1

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75

150

1.00%

1.5

0.50%

0.75

50

1.50%

0.75

0.80%

0.4

Fixed Volume Pipettes

Volume Range

Increment

Test Volume

(μl)

Maximume Permissible Systematic error

   (Inaccuracy)

Maximume Permissible Random error

(Inaccuracy)




%

μl

%

μl

5μl

-

5μl

1.3%

0.065

1.2%

0.06

10μl

-

10μl

0.8%

0.08

0.8%

0.08

20μl

-

20μl

0.6%

0.12

0.5%

0.1

25μl

-

25μl

0.5%

0.125

0.3%

0.075

50μl

-

50μl

0.5%

0.25

0.3%

0.15

100μl

-

100μl

0.5%

0.5

0.3%

0.3

200μl

-

200μl

0.4%

0.8

0.2%

0.4

250μl

--

250μl

0.4%

1.0

0.2%

0.5

500μl

-

500μl

0.3%

1.5

0.2%

1.0

1000μl

-

1000μl

0.3%

3.0

0.2%

2.0

2000μl

-

2000μl

0.3%

6.0

0.15%

3.0

5000μl

-

5000μl

0.3%

15

0.15%

7.5


 



Download:  107-108 Pipettes.jpg109-110 Pipettes.jpg


107-108 Pipettes109-110 Pipettes