Measurement Principle: Zirconia
Display Mode: 320×240 dot matrix color LCD
Measurement Range: 0~10/100/1000PPm/1.00%/25.00% O2
Measurement Accuracy: 0~1000PPm/1.00%/25.00%≤±1%FS
0~100PPm≤±2%FS
0~10PPm≤±5%FS
Resolution: 0.1PPm (trace)
0.01% (normal)
Repeatability: ≤±1.0%FS
Response Time: T90≤30S
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
0-10V.DC (Non-isolated output, load resistance>10KΩ)
2 ways of programmable alarm output (Dry contact type, maximum load 220V/2A)
1 way of error alarm output
Communication Mode: RS232 (default) or RS485
Power Supply: AC88~264V 50/60Hz, power dissipation<25VA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Temperature: 0℃~60℃
Sampling Mode: Gas in
Sample Gas Flow: 400~600mL/min
Gas Inlet Pressure: 5Kpa≤ relative pressure≤200Kpa (relief valve is recommended if the pressure >200Kpa)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or other inert gas
Size: 145mm×175mm×301mm (H×W×D)
Hole-Cutting Size: 135mm×135mm (H×W)
Sensor Service Life: >5 years (under normal conditions of use)
Gas Circuit Port: NPT 1/8 internal thread
Instrument Weight: Net weight 3.1Kg
Installation Method: Embedded mounting