Measurement Principle: Thin-film capacitor
Display Mode: 320×240 dot matrix color LCD, display in ℃ or PPM (selectable)
Measurement Range: Basic: -60℃~+20℃
Standard: -80℃~+20℃
Expanded: -100℃~+20℃
Measurement Accuracy: ≤±1℃(0~-60℃)
≤±2℃ (under -60℃)
Resolution: 0.1℃
Repeatability: ≤±1.5℃FS
Stability: ≤±1.0%FS
Response Time 63%: -60℃→-20℃ 5s
-20℃→-60℃ 45s (pressure 0.1MPa, temperature +20℃)
Response Time 90%: -60℃→-20℃ 15s
-20℃→-60℃ 10min (pressure 0.1MPa, temperature +20℃)
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
0-10V.DC(Non-isolated output, load resistance>10KΩ)
2 ways of programmable alarm output (Dry contact type, maximum load 220V/2A)
1 way of error alarm output
Communication Mode: RS232 (default) or RS485
Power Supply: AC88~264V 50/60Hz, power dissipation <25VA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Dew Point:
Sampling Mode: Gas in
Sample Gas flow: 2~4L/min
Gas Inlet Pressure: 10Kpa≤ relative pressure≤600Kpa
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas:H2, mixed gas of inert gas, hydrocarbon and so on
Size: 145mm×175mm×311mm(H×W×D)
Hole-Cutting Size: 135mm×135mm (H×W)
Sensor Service Life: >4 years (under normal conditions of use)
Gas Circuit Port: NPT 1/8 internal thread
Instrument Weight: Net weight 3.0Kg
Installation Method: Embedded mounting