Measurement Principle: Zirconia
Display Mode: 5 digit high-light LED display
Measurement Range: 10/100/1000PPm~21.000% O2/N2
Measurement Accuracy:10PPm~21.00%≤±1%FS
10PPm~1.00%≤±2%FS
10PPm~100PPm≤±5%FS
Repeatability: ≤±1.0%FS
Response Time: T90≤30S
Output Port: 4-20mA.DC (Non-isolated output, load resistance<1KΩ)
1 way of programmable alarm output (Dry contact type, maximum load 220VAC/2A)
Power Supply: AC85~264V 50/60Hz
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <99%RH, non-condensing
Sample Gas Temperature: 0℃~60℃
Sampling Mode: Gas in
Gas Inlet Pressure: Stable pressure (control the pressure within the recommended limits to make sure the flow is within 400~600mL/min)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: Mixed gas of nitrogen and oxygen
Size: 80mm×160mm×142mm(H×W×D)
Hole-Cutting Size: 76mm×151mm(H×W)
Sensor Service Life: >5 years (under normal conditions of use)
Gas Circuit Port: Φ6 plastic tube fitting
Installation Method: Instrument: embedded mounting
Oxygen sensor: fixing clamp