Measurement Principle: Fuel cell
Display Mode: 128×64 dot matrix display LCD
Measurement Range: 0~10/100/1000PPm O2 (trace)
0~1%/10%/25%/99% O2 (normal)
Measurement Accuracy: 0~1000PPm/1%/10%/25%/99%≤±1%FS
0~100PPm≤±2%FS
0~10PPm≤±5%FS
Repeatability: ≤±1.0%FS
Response Time: T90≤60S (trace))
T90≤40S (normal)
Output Port: Charging port
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <99%RH, non-condensing
Sample Gas Temperature: 0℃~40℃
Sampling Mode: Gas in
Sample Gas Flow: 400~600mL/min
Gas Inlet Pressure: 5Kpa≤ relative pressure≤ 200Kpa (relief valve is recommended if the pressure >200Kpa)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or other inert gas and reducing gas
Explosion-Proof Level: Exib IIC T4
Certificate Number: CNEx16.3613
Gas Circuit Port: NPT 1/8 internal thread
Battery Capacity: 4.2V, 4400mA
Charging Power: AC220V±10% 50/60Hz
Sensor Service Life: >36 months (under normal conditions of use)
Instrument Weight: 3.6Kg
Installation Method: Portable