Measurement Principle: Thin-film capacitor
Display Mode: 64×48 dot matrix OLED display
Measurement Range: Basic: -60℃~+20℃
Standard: -80℃~+20℃
Expanded: -100℃~+20℃
Measurement Accuracy: ≤±1℃(0~-60℃)
≤±2℃ (under -60℃)
Repeatability: ≤±1.5℃FS
Stability: ≤±1.0%FS
Response Time 63%: -80℃→-30℃ 20s
-30℃→-70℃ 5min (pressure 0.1MPa, temperature +20℃, 4L/min)
Response Time 90%:-80℃→-30℃ 40s
-30℃→-70℃ 20min (pressure 0.1MPa, temperature +20℃, 4L/min)
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
Communication Mode: RS485
Power Supply: DC24V±10%
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Dew Point:
Background Gas: mixed gas of H2, inert gas, hydrocarbon, etc.
Sensor Service Life: >4 years (under normal conditions of use)
Installation Method: Instrument: wall mounting, probe: G1/2"screw mounting