Measurement Principle: Fuel cell
Display Mode: 320×240 dot matrix color LCD
Measurement Range: 0~10/100/1000PPm O2 (trace)
0~10%/21%/25%/99% O2 (normal)
Measurement Accuracy: 0~1000PPm/10%/21%/25%/99%≤±1%FS
0~100PPm≤±2%FS
0~10PPm≤±5%FS
Minimum Resolution: 0.1PPm (trace)
0.01% (normal)
Repeatability: ≤±1.0%FS
Response Time: T90≤60S (trace)
T90≤40S (normal)
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
0-10V.DC (Non-isolated output, load resistance>10KΩ)
1 way of programmable alarm output (Dry contact type, maximum load 220V/2A)
2 ways of instrument status signal output
Communication Mode: RS232 (default) or RS485
Power Supply: AC85~264V 50/60Hz, power dissipation<25VA
Environment Temperature: --10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Temperature: 0℃~40℃
Sampling Mode: Gas in or extraction sampling (please indicate sampling mode when ordering)
Sample Gas Flow: 400~600mL/min
Gas Inlet Pressure: Gas in: 5Kpa≤ relative pressure≤100Kpa (relief valve is recommended if the pressure >100Kpa)
Extraction sampling: -10Kpa≤ relative pressure ≤5Kpa(please indicate the specific negative pressure value when ordering if the pressure<5Kpa since it will decide the gas circuit structure)
Gas Outlet Pressure: Free venting (under safe conditions)
Background Gas: N2 or inert, acid and reducing gas, etc.
Size: 176mm×251mm×302mm (H×W×D)
Hole-Cutting Size: 178mm×212mm (H×W) (in embedded mounting)
Sensor Service Life: >24 months (under normal conditions of use)
Gas Circuit Port: NPT 1/8 internal thread
Instrument Weight: Net weight 5.0Kg
Installation Method: Embedded mounting / counter mounting