Measurement Principle: Thin-film capacitor
Display Mode: Segment code display
Measurement Range: Basic: -60℃~+20℃
Standard: -80℃~+20℃
Expanded: -100℃~+20℃
Measurement Accuracy: ≤±1℃(0~-60℃)
≤±2℃ (under -60℃)
Resolution: 0.1℃
Repeatability: ≤±1.5℃FS
Stability: ≤±1.0%FS
Response Time 63%: -60℃→-20℃ 5s
-20℃→-60℃ 45s (pressure 0.1MPa, temperature +20℃)
Response Time 90%: -60℃→-20℃ 15s
-20℃→-60℃ 10min (pressure 0.1MPa, temperature +20℃)
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
1 way of programmable alarm output (Dry contact type, maximum load 220V/2A)
Power Supply: DC24V±10%, 200mA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Dew Point:
Sampling Mode: Diffusion/ Gas in (please indicate the sampling mode when ordering)
Sample Gas Pressure: Diffusion (normal pressure dew point): stable pressure
Gas in (normal pressure dew point: relative pressure ±10kPa, stable pressure (control the pressure within the recommended limits to make sure the flow is within 2~4L/min)
Background Gas: Inert gas and non-corrosive gas such as hydrocarbon
Size: 206mm×135mm×97mm (H×W×D)
Explosion-Proof Leve: Exd IIC T6 Gb
Certificate Number: CNEx20.1001
Protection Level: IP65
Sensor Service Life: >4 years (under normal conditions of use)
Instrument Weight: Diffusion: Net weight 2.0Kg
Gas in: Net weight 2.4Kg
Cable Connection: 3 meters (standard)
Installation Method: Screw mounting (diffusion)/ bracket mounting (gas in)