Measurement Principle: Fuel cell/ thermal conductance (optional)
Display Mode: 320×240 dot matrix color LCD
Measurement Range: 0~1%/30%/50%/99.99% H2 (thermal conductance)
0~10/100/2000PPm H2 (fuel cell)
Measurement Accuracy: ≤±1.0%FS (thermal conductance)
≤±1.5%FS (fuel cell)
Resolution: 0.01% (thermal conductance)
1PPm (fuel cell)
Repeatability: ≤±1.0%FS
Response Time: T90≤60S
Output Port: 4-20mA (Non-isolated output, load resistance<1KΩ)
0-10V.DC(Non-isolated output, load resistance>10KΩ)
2 ways of programmable alarm output (Dry contact type, maximum load 220V/2A)
1 way of error alarm output
Communication Mode: RS232 (default) or RS485
Power Supply: AC88~264V 50/60Hz, power dissipation <25VA
Environment Temperature: -10℃~+50℃
Storage Environment Humidity: <90%RH, non-condensing
Operation Environment Humidity: <100%RH, non-condensing
Sample Gas Temperature: 0℃~50℃
Sampling Mode: Gas in
Sample Gas Flow: 400~600mL/min
Gas Inlet Pressure: Thermal conductance: 5Kpa≤ relative pressure≤200Kpa (relief valve is recommended if the pressure >200Kpa)
Fuel cell: 5Kpa≤ relative pressure≤100Kpa (relief valve is recommended if the pressure >100Kpa)
Gas Outlet Pressure: Free venting (under safe conditions)
Size: 145mm×175mm×301mm (H×W×D)
Installation Hole Size: 135mm×135mm (H×W)
Sensor Service Life: Thermal conductance >5 years, fuel cell>3 years (under normal conditions of use)
Gas Circuit Port: NPT 1/8 internal thread
Instrument Weight: Net weight 2.8Kg
Installation Method: Embedded mounting