Industrial Sludge Concentration Sensor Output 4-20mA

Model
Nature of the Manufacturer
Manufacturer
Product Category
Turbidity&TSS(MLSS)
Place of Origin
City, Province
Overview
★ Model No: TCS-1000/TS-MX★ Outputl: 4-20mA★ Power Supply: DC12V★ Features: Scattered light principle, automatic cleaning system★ Application: power plant, pure water plants, sewage treatment plants, beverage plants,environmental protection departments, industrial water etc
Product Details

Introduction

Online Suspended solid sensors for on-line measurement of scattered light suspended in the degree of opaque liquid insoluble particulate matter produced

by the body and can quantify levels of suspended particulate matter. Can be widely used in site online turbidity measurements, the power plant, pure water

plants, sewage treatment plants, beverage plants, environmental protection departments, industrial water, wine industry and the pharmaceutical industry,

epidemic prevention departments, hospitals and other departments.

Features

1. Check and clean window every month, with automatic cleaning brush, brush a half hour.

2. Adopt sapphire glass realize easy maintain, when cleaning adopt scratch-resistant sapphire glass, don’t worry about the wear surface of window.

3. Compact, not fussy installation place, just put in to can complete the installation.

4. Continuous measurement can be achieved, built-in 4~20mA analog output, can transmit data to the various machine according to the need.

Technical Indexes

Model No.TCS-1000/TS-MX
Measuring range0-50000mg/L(kaolin)
Power supplyDC24V±10%
Current drawAt regular operation: 50mA (Max.), At cleaning operation: 240mA (Max.) (excluding analog signal output)
OutputAnalog (4-20mA) signal output: Resistance load of 300Q(Max.)

Self-checking output: open collector (DC24V 20mA Max.)

InputCalibration signal input
Cleaning systemAutomatic wiper cleaning system
Time interval for cleaningClean once immediately after power-on, and subsequently clean once every 10 minutes
Operating temperature0 to 40°C (unfrozen)
Major materialSUS316L, Sapphire glass, Fluorocarbon rubber, EPDM, PVC (cable)
Dimensions48x146mm
WeightApprox. 1.1kg
Degree of protectionIP68, Maximam depth of 2m(underwater type)
Detector cable length9m

What is Total Suspended Solids(TSS)?

Total suspended solids, as a measurement of mass are reported in milligrams of solids per liter of water (mg/L) 18. Suspended sediment is also measured in mg/L 36. The most accurate method of determining TSS is by filtering and weighing a water sample 44. This is often time consuming and difficult to measure accurately due to the precision required and the potential for error due to the fiber filter 44.

Solids in water are either in true solution or suspended. Suspended solids remain in suspension because they are so small and light. Turbulence resulting from wind and wave action in impounded water, or the movement of flowing water helps maintain particles in suspension. When turbulence decreases, coarse solids quickly settle from water. Very small particles, however, may have colloidal properties, and can remain in suspension for long periods even in completely still water.

The distinction between suspended and dissolved solids is somewhat arbitrary. For practical purposes, filtration of water through a glass fiber filter with openings of 2 μ is the conventional way of separating dissolved and suspended solids. Dissolved solids pass through the filter, while suspended solids remain on the filter.