Application:
The DP Series vacuum ovens provide the right temperature and vacuum environment. Do vacuum drying or thermal test under vacuum conditions in the fields of electrical, electronics, aviation and other related fields. The vacuum drying oven can independently perform low pressure test, or comprehensive high temperature for comprehensive test. Based on years of experience in the field of environmental simulation equipment, the company has optimized the temperature field of the vacuum drying oven, and the technical indicators have reached the first-class level in the industry.
Features:
1. The shell of the vacuum oven is made of cold rolled steel plate and coated with double sided electrostatic epoxy polyester powder.
2. The pressure-bearing inner chamber is made of SUS304 stainless steel plate.
3. Use centrifugal glass insulation cotton felt insulation. The hinge of working chamber is installed on the left side of the oven to ensure optimum contact pressure when the door is closed.
4. The controller is installed in the middle or lower part of the oven.
5. The electrical components are installed in the electric control cabinet, and the operator and test samples can be fully protected.
Function principle:
1. The door of working chamber is made of silicone rubber seals, to ensure a good seal at low air pressure.
2. The low-pressure environment inside the oven is realized by the vacuum pump pumping the air inside the oven out of the oven.
3. The temperature environment inside the oven is provided by heater installed outside the pressure-bearing inner chamber.
4. Use Japanese Yokogawa UT155 or Korean Autonics controller to control the temperature and pressure.
5. A separate set of temperature sensors is also provided to protect the test chamber and the testing pieces.
6. Pressure can be restored automatically or manually.
Standard configuration:
1. Internal pressure-bearing external insulation box: 27-216L;
2. The air pressure can be as low as 1 kPa;
3. Nickel-chromium alloy electric heater;
4. PID+SSR temperature control;
5. Hysteresis pressure control;
6. Independent over-temperature protection system;
7. The door opening angle is 120°;
8. Observation window (003,005: no; 01, 02: 400W*400H mm);
9. Shelf: 2 pieces;
10. Back wall Φ33 blind flange;
11. 4 casters;
Optional accessories:
Oil mist trap, cold trap, other options and special requirements can be added according to user needs;
Note:
1. The laboratory is sensitive to oil mist, oil mist traps can be used. You can also connect the exhaust pipe by yourself and remove the oil mist from the laboratory.
2. The test sample contains more water, and the vacuum pump needs to be changed to the type with gas ballasting. When drying, open the gas ballasting.
3. When drying the samples, organic matter such as oil mist evaporates, then can use a cold trap, to adsorb organic matter onto the surface of the cold trap, in order to avoid organic matter into the vacuum pump lubricant.
Technical parameter:
Model | HD-01DP | HD-02DP |
Volume | 90L | 216 L |
Internal dimensions (D*W*H) | 450*450*450 mm | 600*600*600 mm |
External dimensions (D*W*H) | 670*670*1520 mm | 820*820*1670 mm |
blind flange | Φ 33 mm, on the back wall of the drying oven |
Performance |
Temperature range | Room temperature + 20℃...200℃ |
Temperature fluctuation | ≤ 1 ℃ |
Pressure range | Atmospheric pressure... 1kPa |
Controller | International well-known brand temperature controller, digital display temperature control, with PID self-tuning function, PID+SSR temperature control |
Heater | Nickel-chromium alloy electric heater |
Power supply | AC 380V/50Hz, three-phase four-wire + earth wire |
Power | 2.5 kW | 3.5 kW |
Noise | ≤ 70 dB(А) |