Force, Accelerometer, Displacement, Pressure, Angular Rate, Voltage, Current Sensors Calibration System.- According to JJF, GB/T, ISO, SAE, etc. Regulation;
- Static Calibration;
- Calibration Range by Required;
- Calibration Fixture Supplied;
- Calibration Parameters by Required.
SCS-XXX serials calibration system mainly focus on vehicle safety test field, including dummy loadcell, seatbelt loadcell, pedestrian protection legform loadcell, dummy and vehicle body accelerometers, kinds of displacement (as IR-TRACC), Gyro, door pressure sensor, airbag pressure sensor, airbag squib current sensors, and EV/HEV high voltage and current sensor. These calibration systems are suitable to the sensors of general industry.
Specification (Force Calibration System SCS-FO):
| Name | Unit | Value |
| Range | KN | ≤45 |
| Loading Method | / | Step/Continues |
| Cal. Index | Sensitivity, offset, Non-Linearity, Hysteresis, Crosstalk, Shunt. |
| Excitation | V | 2.5/3.3/5/10 |
| Seatbelt Loadcell Cal. | ISO 17242 | Support |
| Sensor Test Chain | Amp≤10V | Support |
Specification (Accelerometer Calibration System SCS-AC):
| Name | Unit | Value |
| Amp. Cal. Range | g | 10~1000 |
| Freq. Cal. Range | Hz | 10~10000 |
| Regulation | ISO16063 | GB/T20485 |
| Cal. Index | Sensitivity, Deviation, Offset |
| Excitation for PR | V | 5/10 |
| Excitation for IEPE | mA | 4 |
| PE Accelerometer | pC input | Support |
| Sensor with Test Chain | Amp≤10V | Support |
Specification (Displacement Calibration System SCS-DS):
| Name | Unit | Value |
| Cal. Range | mm | <2000 |
| Cal. Objects | H3/ES2/WS/THRO/Q Serial Dummy Rib Displacement, Knee Pot. and Other String Pot. |
| Cal. Index | Sensitivity, Non-Linearity |
Specification (Pressure Calibration System SCS-PR):
| Name | Unit | Value |
| Cal. Range | kPa | ≤600 |
| Loading Method | Static | JJF1640-2017 |
| Cal. Index | Sensitivity. Offset. Non-Lin. Hys. |
| Excitation | V | 5/10 |
Specification (Angular Rate Calibration System SCS-AV):
| Name | Unit | Value |
| Cal. Range | deg/sec | ≤18000 |
| Loading Method | Static | JJF1535-2015 |
| Cal. Index | Sensitivity. Offset. Non-Lin. Hys. |
| Excitation | V | 5/10 |
Specification (Voltage Calibration System SCS-VO):
| Name | Unit | Value |
| Cal. Range | V | ≤1000 |
| Loading Method | Static | DAQ by Step |
| Cal. Index | Sensitivity. Offset. Non-Lin. Hys. |
| Excitation | V | 5/10 |
| 3-Ch Voltage Sensor | Optional for EV/HEV V1,V2,Vb |
Specification (Current Calibration System SCS-CU):
| Name | Unit | Value |
| Cal. Range | A | ≤2500 |
| Cal. Index | Sensitivity. Offset. Non-Lin. Hys. |
| Excitation/Power Supply | V | 5/10/±15/24 |
Note: Ampere-turn Method applied Above 100A.