- Back-to-Back Calibration Installing;
- According to ISO16063-21;

- Support PR, IEPE, DIR Sensor Types;
- Measure Offset, Bias Voltage, Bridge Resistance;
- Frequency Range 20Hz~10kHz;
- Vibration Max. Moving 8mm;
- Transverse Moving≤5%;
- Sampling Frequency≥500kHz, 16bit A/D;
- Force Rating 45N Sine Peak;
- Peak≥20g@150g Payload; Max. Payload 500g;
- Expanded Uncertainty (k=2): ±1.0% (80Hz~160Hz); ±1.5% (20Hz~2500Hz); ±2.5% (2500Hz~10kHz).
Accelerometer Calibration System SCS-AC, according to ISO16063-21, is very suitable to middle-high frequency range (20Hz~10kHz). System include a vibration exciter BK4809, which is very suitable to small object. More force rating type BK4808 is optional if needed. The standard accelerometer Endevco 2270 or 910A-500 is good at calibration application (Resonant Frequency≥35kHz). System control and report generation are by software automatically.
Technical Specification (22℃±2℃, 30%RH~75%RH):
Technical Specification (22℃±2℃, 30%RH~75%RH):
| Name | Unit | Value |
| Frequency Range | Hz | 20~10000 |
| Force Rating | N | 45 |
| Max. Moving | mm | 8 |
| Calibration Result | Sensitivity, Deviation, Offset… | |
| PR Excitation | V | 5 or 10 |
| PR Signal | mV | ±1000 |
| IEPE Constant Current | mA | 4 |
| IEPE Max. Voltage | V | 21 |
| IEPE Signal | V | ±10 |
| DIR Signal | V | ±10 |
| Aided Power Output | V | 24,±12 |
| Vibration Exciter | B&K 4809,Max. Acc 75g Max. Velocity 1.65m/s Max. Driver Current 5A RMS Operation Temp. 5℃~50℃ | |
| Power Amplifier | B&K 2718,Max. 75VA; Output Max. Current 5A RMS 10Hz~20kHz Response Error <±0.5dB | |
| Name | Unit | Value |
| Sensitivity | pC/g | 2.2 |
| Response (-3dB) | Hz | 2~20000 |
| Transverse Sensitivity | <3% | |
| Amplitude Error | per 1000g | <0.1% |
| Sensitivity Stability | <±0.2% / year | |
| Name | Unit | Value |
| Signal Input Range | V | ±10 |
| Sampling Frequency | Hz | 500k |
| Resolution | 16bit; Gain Error<200ppm | |
| Input Impedance | Ω | >100G |

![]() | Accelerometer Calibration System SCS-AC |
,