PGC-2020C VOCs online monitoring system

Model
Nature of the Manufacturer
Other
Product Category
VOCS Online Monitoring System
Place of Origin
City, Province
Overview
PGC-2020C VOCs online monitoring system
Product Details

System introduction:

PGC-2020c environmental air non methane hydrocarbon on-line monitoring system collect the environment sample by negative pressure generated by an vacuum pump. The samples pass through the atmospheric sampling header to the NMHC gas chromatograph. With FID method to monitor the non methane hydrocarbon content in the environment is monitored on-line.


System component:

Pgc-2020c ambient air non methane hydrocarbon is consist of atmospheric sampling pipe and 19 inch standard machine

Cabinet, NMHC gas chromatograph, online air generator, online hydrogen generator, zero level air purifier, industrial computer, dynamic diluter and special configuration software.



System parameter:

Automatic sampling structure design to sure automatic continuous monitoring on-line.


Large LCD screen is used to display the data.

Professional system software, powerful, could display visual of various test data and tendency map, automatic calculation of daily average, monthly average, pollution index, etc.,

And generate variety of reports and save in the system, and can be printed, etc; 

gas chromatographs adopt EPC to control gas flow, and can be equipped with different high sensitivity detectors; 

Multiple exchange control interfaces (4-20mA, RS485, RS232, LAN); 

Easy installation and good anti-interference ability.



technical parameter:

output

Air requirement

High purity nitrogen or zero polar air (≥99.999%); Combustion gas: high purity hydrogen (≥99.999%); auxiliary gas: zero grade air (hydrocarbon < <20PPb)

Operation environment condition

(-10-50)℃,Humidity:(10%-90%)RH

Power supply

<5000W,220VAC/50Hz

Analyze duration

15~30min

Range drift

±1%Full range

Zero drift

Zero drift free

Repeatability

RSD≤3%

Detect limit

≤0.03PPb(normal butane),≤0.02PPb(benzene)

detector

Hydrogen flame ionization detector (FID)

range

0.01PPb-1ppm(benzene)

Test capacity

VOCs(C2-C12)

project


        4~20mA,RS232/RS485,ethernet network


Petrochemical, chemical, pharmaceutical, building materials, spraying, etc., and large industrial and mining enterprises discharge outlets