Welcome Customer !

Membership

DongGuan HongTuo Instruments Co., Ltd.
Free Member ·1YRS
,
Other

Main Products: 检测设备

Xwboo>Products

Methyl Silicone Oil Plastic Testing Machine For Heat Deflection Temperature And Vicat Softening T...

Model
Nature of the Manufacturer
Other
Product Category
Plastic Testing Machine
Place of Origin
City, Province
Overview
Place of Origin: Dongguan China Brand Name: DahoMeter Certification: ISO,SGS ......
Product Details
Product Name: HDT Vicat Testing Machine Heating Medium: Methyl Silicone Oil
Cooling Method: Water Cooling Package: Safe
Power: Electronic Testing Before Delivery: Yes
High Light:

plastic testing instruments

,

plastic pipe testing equipment

Plastic Testing Machines for Heat Deflection Temperature and Vicat Softening Temperature


The thermal deformation, Vicat softening point temperature tester is mainly used for the determination of temperature of thermal deformation and Vicat softening point to nonmetallic materials(such as plastic, rubber, nylon, electric insulating materials, etc.). It is the ideal testing tool for chemical industry enterprises, research institutes, colleges and so on. It adopt computer controlled.

Standard

GB/T 1633, GB/T 8802, GB/T 1634, ISO 2507, ISO 75, ISO 306, ASTMD 1525, ASTMD 648, BS 2782 Part 1 1970, DIN 53460

Features

1. Computer collects temperature fluctuation and controls overall process, real time displays temperature, deformation curve and tes time, CRT color curve display, prints the test result automatically.

2. The software adopts menu operation, which makes operation of users more practical.

3. Digital process, computer display real-time temperature, deformation and test time at the same time, which makes the system more reliable.
4. Report forms can be automatically compiled and meet with the requirements of customer.

5. Sample frame can be lift and drop automatically.

6. Adopt dial gauge to measure deformation, more accurate than adopting displacement sensor.

7. Multi safety protection, mechanical over temperature protection, software over temperature protection, auto stop after test complete, user password protection.

8. Adopt advance PID technology and photo electric technology to control temperature raise, ensure error of heating speed is less than 0.5C.

Technical Parameter

Temperature Control RangeRoom Temperature-300C
Max Temperature Error±0.5C
Heating Rate50C/h-120C/h (adjustable)
Test Station3
Deformation Measurement MethodHigh Precision Dial Gauge
Deformation Measurement Range0-10mm
Max Deformation Error+/-0.005mm
Resolution of Deformation0.001mm
Sample Span64mm, 100mm
Thermal Expansion Coefficient of Sample Frame0.005/100C(important)
Sample FrameAutomatic Lift and Drop
Heating MediumMethyl silicon oil
Cooling MethodNatural Cooling or water cooling below 150C, Natural Cooling Above 150C
Dimension950mmX660mmX1200mm
Weight230 Kgs
Power Supply220V, 50Hz, Single Phase

Similar Product Recommend