Short Description: The system adopts innovative liquid circulation refrigeration combined with Peltier temperature control technology.Light and small equipment size meets the on-site instantaneous detection demands.The system supports the application of all common real-time fluorescent quantitative PCR detection modes Model: QPCR-KP-Archimed Mini 16 Noise: Temp Accuracy: ±0.2ºC ......
Product Introduction
In order to meet the needs of on-site rapid and instantaneous detection such as epidemic prevention and control, entry-exit rapid screening, food safety, environmental microorganism identification and small flux scientific research and teaching, the ultra-fast and portable fluorescence quantitative PCR system ArchimedTM Mini 16 was launched. The system adopts innovative liquid circulation refrigeration combined with Peltier temperature control technology to ensure getting results quickly and accurately. Light and small equipment size, which can be free from the constraints of the site, meets the on-site instantaneous detection demands. The system supports the application of all common real-time fluorescent quantitative PCR detection modes, including qualitative detection, relative quantification, absolute quantification, genotyping, etc.
Product Features

Technical Specifications
Thermal control | Optical Detection |
Sample Capacity | 16 (0.2mL PCR tube or strips) | Light Source | 2 or 4 monochrome efficient LEDs |
Reaction Volume | 10-50 uL | Optical Detector | High-performance MPPC |
Thermal Cycling Tech | Peltier | Detection Mode | Ultra-fast Time-resolved Scanning |
Cooling Method | Liquid cycle refrigeration | Fluorescent Channel | FAMSYBR Green, VICJOE/HEX/TET, JUN, ROX/Texas Red (optional) Mustang Purple, Cy5/L1Z (optional) |
Maximum Ramp Rate | 8℃/s | Running Time | <30 minutes (combined with rapid kit) |
Temp Accuracy | ±0.2℃ | Sensitivity | Single copy of the gene, the difference of 1.33 times concentration can be distinquished |
Temp Homogeneity | ±0.2℃ @60℃ ±0.2℃ @95℃ | Dynamic Range | 10 orders of magnitude |
Other Configurations |
Instrument Noise | <50 dB | Size | 205*190*98 mm (L*W*H) |
Analysis Module | 220VAC, 50/60hz | Operation System | Windows |
Analysis Module |
Qualitative Detection, Absolute Quantification, Relative Quantification, Genotyping |