Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler

Model
Nature of the Manufacturer
Other
Product Category
Dissolution Tester
Place of Origin
City, Province
Overview
Short Description: The sampling rack is automatically raised and lowered.Solvent lighting (including dark light drug with red light)Automatic centering cup holder and cup holder ringThe water bath and circulation line are connected. Model: QYXT-RSD-RZQ-12K Maximum times of sampling: 15
Product Details

Introduction

The QY-8K/12K Dissolution Sampling System is a new instrument designed in accordance with the requirements of the Chinese Pharmacopoeia and the United States Pharmacopoeia. They are patented products, test instruments that meet the quality consistency evaluation of generic drugs.

Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler Biometer 12 Cups Intelligent Dissolution Sampling System Integrated Automatic Sampler

Main Features of QYQ-RSD-QY-12K

●High resolution color touch screen

●Built-in printer

●Fingerprint recognition

●Using precision glass syringe as sampling power

●Using 12 channel high precision, high chemical stability valve body

●High chemical stability PTFE pipeline has excellent anti-adsorption function

●Filter infiltration technology before test to reduce adsorption

●Sampler’s liquid return circulation technology and pipe emptying technology after sampling avoid cross contamination

●Has the same amount of isothermal rehydration function

●With distribution, dilution function

●Has 12 channels of independent piping system to meet the requirements of the dissolution test for blanks and chemical reference standard

●Sampling double-stage filtration, primary filtration is a columnar filter head with a pore size of 10 µm, and secondary filtration can choose a ɸ25mm injection filter with 0.8 µm/0.45 µm/0.22 µm pore diameter or no filtration

●15-bit test tube rack for up to 120 test tubes

●Sampling tubes with a silicone cap to prevent evaporation and contamination of the sample

●Color touch screen, Chinese menu prompts, easy to operate

●Can be run online with the dissolution apparatus

●Ampoule collection device (optional)

Parameters of QYQ-RSD-QY-12K

Number of sampling channels12
Maximum times of sampling15
Maximum allocations15
Longest sampling interval99 hours 59 minutes
Minimum sampling interval1 minute for the first time, 2-10 minutes for the remaining (depending on the sampling volume and whether it is rehydration, distribution, dilution)
Continuous working hours99 hours 59 minutes *15
Sampling filter period<30 seconds
Sampling volume (or fluid infusion volume)1ml-100ml
Single tube distribution amount0-15ml
Single tube dilution0-15ml
Sampling accuracyRSD1% (average of each channel when the sample volume is 10ml)

Main Features of RCY-RSD-RC-12K

1. Equipped with 14 vessels, 12 cups and 12 paddles, it can meet the requirements of the control test for blank and chemical reference standard.

2. Head electric lift, high degree of automation.

3. Standard spraying method and cup method are available. Three kinds of dissolution test methods such as small cup method, dish method (album, small dish) and drum method (long tube, short tube) are available.

4. The automatic height-adjusting clutch is used, and it is not necessary to manually set the height of the rotating rod. Different test methods are applicable, and it is not necessary to repeat the height after replacing different test devices.

5. With automatic drug delivery function, can achieve simultaneous drug delivery and sequence administration.

6. Can be administered by granules.

7. The sampling rack is automatically raised and lowered to automatically determine the sampling point.

8. 14-channel solvent temperature monitoring, can be set to real-time monitoring or sampling monitoring according to user habits.

9. Solvent lighting (including dark light drug with red light)

10. Built-in WiFi module

11. Can be run online with the sampler.

12. The bottom surface of the head is designed with a tapered sealing cup cover, so that the evaporation droplets of the solvent fall back into the dissolution cup to avoid contamination caused by evaporation of the solvent.

13. Adopts automatic centering cup holder and cup holder ring, and is equipped with high-precision dissolution cup. Realize the automatic centering of the rotating rod and the dissolution cup without manual adjustment.

14. The water bath and circulation line are connected by quick joints and have independent water discharge ports for easy disassembly and cleaning.

15. LCD touch screen display, intuitive operation interface, easy to use.

16. Has a user management function, which can be logged into the user account by means of a user name plus a password and a fingerprint, and has a quest account, and any user can log in to the account after being enabled.After logging in, various parameter settings and running tests can be performed under this account.

17. Speed regulation can be in different dissolution instrument zone, which can do different speed comparison test at the same time.

Parameters of RCY-RSD-RC-12K

Stirring paddle swing amplitude≤0.5mm
Cup swing amplitude ≤1.0mm
Deviation between the rotating rod and the vessel axis≤2.0mm
Speed range(10-300) rev / min
Steady speed error≤±1rpm
Temperature adjustment rangeRoom temperature -45.0°C
Temperature resolution0.01°C
Temperature control error≤0.2°C / ±0.3°C
14 cups of solvent temperature monitoring resolution0.01°C
Dosing interval 10 seconds – 999 seconds
Sample point resolution1ml
Basic test timing≤999hours 59 minutes
Program-controlled test program99 groups * 7 users
The times of program-controlled sampling15 times. Meet the requirements of quality consistency evaluation of genetic drugs.
Program-controlled sampling time≤999hours 59 minutes
Working noise<60dB (A)
Working environmentAmbient temperature (5-35)°C, relative humidity ≤80%
Safety performanceIn line with national industry standards
Power requirements AC 220V 50Hz 10A