Short Description: The disintegration tester adopts a single-chip microcomputer control system.The temperature of the water bath is controlled by an integrated temperature sensor.It can send out sound and light warning signals, and has an automatic protection functionThe lifting movement time of the two groups of hanging baskets is separately controlled by two synchronous motors. Model: BJY-RSD-KB-1G Power supply: AC 220V......
Introduction
The KB-1G Intelligent Oral Disintegration Tester is a mechatronic drug test instrument developed in accordance with the provisions of the Pharmacopoeia of the People’s Republic of China (2015 Edition) on the disintegration time limit testing of tablets and pills. Its main technical indicators also comply with the American U.S.P-23 version, the British BP-1988 version and the 11th edition of the Japanese Pharmacopoeia on disintegration time limit testing.
Features
1. The instrument adopts a single-chip microcomputer control system, and the temperature of the water bath is controlled by an integrated temperature sensor.
2. The lifting movement time of the two groups of hanging baskets is separately controlled by two synchronous motors.
3. When the set testing process is over, or when the testing or control system fails, and when the temperature of the water bath is super high or low, the instrument can send out sound and light warning signals, and has an automatic protection function.
Technical Parameters
Temperature preset range | Freely set between 5.0° C(or room temperature)-40.0° C, display resolution: 0.1° C |
Temperature control accuracy | ≤±0.5° C |
Timing range | freely set within 10 hours, display resolution is in minute |
Lifting frequency of hanging basket | 30 times / min |
Lifting distance of hanging basket | 10mm±1mm |
When the disintegration basket is lowered to the lowest position, the screen is below the water surface | 15mm±1mm |
When the disintegration basket rises to the highest position, the screen is below the water surface | 5mm±1mm |
Screen aperture | 710µm |
Working Conditions
Power supply | AC 220V±10% 50Hz |
Ambient temperature | 5°C-35°C |
Relative humidity | ≤80% |
Power | 550W |