Short Description: Transparent sleeve, stainless steel net racksAutomatic temperature control, high temperature control precisionThree beakers, three-paths, three-path synchronization, automatic overturn.Double-functional head can ascend and descend electrically and turn over manually. Model: RBSXY-JT-RBY-4B Dimension: 520mm*380mm*470mm Timing range: (10~900)min Volume of be......
Product Description
RBY-4B thawing time limit tester is used to inspect the condition of thawing, softening or dissolution of solid preparations such as suppositories or vaginal tablets under the prescribed conditions.
It’s a mechatronic product which is designed and produced in accordance with the requirements of Chinese Pharmacopoeia (Edition 2010) on ”suppository thawing time limit inspection methods”.
Technical Indexes
Number of sleeve racks | 3 sets |
Size of transparent sleeve | Height:60mm; inner diameter: 52mm; each has 39 holes of 4mm |
Timing range | (10~900)min |
Timing precision | ±0.5min |
Temperature control range | (20~45)ºC |
Temperature control precision | ±0.5ºC |
Volume of beaker | 5L(3PCS) |
Power supply | 220V/50Hz/1000W |
Main Features
1. Three beakers, three paths, three-path synchronization, automatic overturn.
2. Double-functional head can ascend and descend electrically and turn over manually.
3. Transparent sleeve, stainless steel net racks
4. Imported stainless steel is adopted for net racks.
5. Automatic temperature control, high temperature control precision
6. Magnetic water pump circulating water flow uniform heating system is adopted, and bath temperature is uniform.
7. Automatic temperature control, automatic time keeping, timing overturn indication forecast and fault alarm.
8. Three parameters including running time, overturn frequency and bath temperature can be preset arbitrarily; preset value and real-time value are displayed by time division.
The system provides three kinds of shortcut modes for customer choice:
P1: run for 30min, the metal racks turn over once automatically every 10 minutes
P2: run for 60min, the metal racks turn over once automatically every 10 minutes
P3: run for 30min, the metal racks don’t turn over.