Welcome Customer !

Membership

Qingdao Jiading Analytical Instruments Co., Ltd.
Free Member ·1YRS
Other

Main Products: 检测仪器

Xwboo>Products

Biometer Automatic LCD Display Test Apparatus Kjeldahl Nitrogen Analyzer

Model
Nature of the Manufacturer
Other
Product Category
Nitrogen Analyzer
Place of Origin
City, Province
Overview
Short Description: Built-in nitrogen content calculatorAuthentic color LCD touch screenRegular nitrogen determining moduleWater-saving design, warning reminder of low water pressure and auto power-off reminder Model: DDY-SS-K1305 Dimension: 420*365*710mm Recovery Rate: ≥99.5 Repeatability: 0.5% Relative Error
Product Details

Product Introduction

It meets international / national standard, and adopts national standard Kjeldahl method: digest sample in concentrated sulphuric acid environment, steam distillation in alkaline environment, boric acid absorption, and indicator titration end-point color determination method.

The distillation procedure could be finished automatically: automatically add water, add alkali and distill, and automatically power-off and give sound reminder when the operation is finished.


Highlighted Features

The authentic color LCD touch screen is equipped for the control of alkali adding and distillation, and the condition can be stored. Once distillation is finished, there will be sound reminder;

Regular nitrogen determining module is equipped, and upgrade interface for auto titration and inducted interface for reagent barrel are reserved;

Water-saving design, warning reminder of low water pressure and auto power-off reminder for water supply cut-off are available;

Pre-heating, water supply, steam pressure control, dry burning prevention, and cleaning can all be realized automatically;

Built-in nitrogen content calculator displays the content of nitrogen or protein once the sample amount and standard titration solution consumption amount are input.


Meticulous Fault Self-diagnosis and Control Function

1. Not-in-position prompt of security door and digestion tube;

2. It is guaranteed that the circulating water flow-limiting valve open only during the distillation, which is save water, free from worry, automatically control circulating water flow; Reserved interface for automatic circulated cooling water machine, apply to dry conditions or water-saving laboratory;

3. Warning reminder of low water pressure and auto power-off reminder for water supply cut-off are available;

4. Mature and stable steam generator technology is applied, automatic preheating, water supplying, steam pressure controlling, dry burning prevention, and cleaning can all be realized automatically;

5. Steam generator is equipped with pressure sensor, temperature sensor. temperature protection switch, water level double detection, and other protections like these;

6. The separated specialized bellows pump is resistance to the erosion and particle plugging. accurate measuring of the added acid and lye, interior is made of PTFE(Teflon). The added reagent is measured by volumetric, no need calibration. long lifetime up to 5-10 years(by end-users’ verification);

7. Distillation system is specially designed according to the rule of ammonia escape, ensure high reliable recovery rate. get unanimous approval from national testing institution, and invited to draft national standard;

8. Double distillation modes are optional;

9. Auto nitrate nitrogen determining module is available, for easy machine operation of the nitrate nitrogen;

10. Upgrade interface for auto titration and inducted interface for reagent barrel are reserved

11. The digestive tube is specially designed surrounded with intelligent facilities, and the transparent acrylic cover is equipped for total protection;

12. The panel is with the material of ABS engineering plastic, which is waterproof, electricity-proof and corrosion-proof.


Technical Parameters

Type

K1305A

K1305B

Function

Regular nitrogen determining module

Both regular nitrogen determining module and nitrate nitrogen determining module

Measuring range

Nitrogen 0.1-240mg

Distilling time

3 to 8 minutes per sample per sample

Regular determination:

3 to 8 minutes per sample per sample

Nitrate nitrogen determination: 10 to 18 minutes per samples

Repeatability

Relative error 0.5%

Setting of steam flow

0-99% (Strengthened version)

Recovery rate

99.5%

Reagent volume

0-150ml, 10ml/level

Setting time

0-99 minutes

Delaying time

0-99 minutes

Deescalating time

0-99 minutes

Rated power

1800W

Dimensions

420*365*710(mm)

Weight

30kg



Similar Product Recommend