Welcome Customer !

Membership

Jinan Xinghua Instruments Co.,Ltd.
Free Member ·1YRS
,
Other

Main Products: 检测设备

Xwboo>Products

XHD-01A Filter integrity tester

Model
Nature of the Manufacturer
Other
Product Category
Pharmaceutical Testing Equipments
Place of Origin
City, Province
Overview
XHD-01A Filter integrity tester
Product Details

Product Description

Integrity Tester are designed for testing integrity of filters and filter systems . The State Pharmacopoeia and GMP specification requirements. The Integrity Tester is compact, handy to use, and fully automated integrity test instrument, which performs bubble point, diffusion flow, enhanced bubble point, and water-based test for hydrophobic filters. The first domestic launch of integrity test for water-based test for hydrophobic filters to meet different customer needs.



Test Standards

Product Features

1.The scientific electronic signature and classification to distinguish the responsibility easily, and to prevent false operation.
2.Display of test data and curves in real time, monitoring the whole process of testing.
3.Provides automatic printing set function, the user can operate more simple and convenient.
4.Water Based test (WH) for Hydrophobic Filter: using purified water as test liquid instead of IPA and Ethanol, thus it can avoid trace quantity of ethanol or IPA contamination when testing filters.
5.The instrument can store 500 history record and curve.


Technical Parameter

Applied Range:

Disc Membrane:Φ25mm-Φ300mm;
Standard Cartridge:2.5″- 40″
Capsule & Mini Catridge
Air filter test :2.5″- 40


Power Supply

100 – 260 VAC, 50HZ, 120W

Operation pressure

100-8000mbar (115psi)

Unit

mbar

Operation Condition

Ambient Temperature:+5℃ ~ +40℃: Relative Humidity:10-80%

Dimension(mm)

400(Depth)*240(Width)*270(Height1)*220(Height2)

Test Function

Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test;Diffusion Flow Test;Water Immersion Test;

Test Accuracy

Upstream Volume:± 4%;Bubble Point:≤± 50mbar ;Diffusion Flow Test:≤± 4% ;Water Immersion Test:≤± 0.05ml

Test Scope

Bubble Point :590-6900mbar Diffusion Flow Speed:1-180ml/min WI:0.01-30ml/min

Applied Range

Symmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (3 core within 20 inches)

Print Function

Built-in printer;

Historical Records

500 sets;

Display Screen

5.7″TFT,Single color;

Serial Connection

RS232;

Language

Chinese/English;

Work Mode

Online/Offline;

Applicable Environment

Above D level;

Weight

7.5KG;



Relative Products

Similar Product Recommend