Welcome Customer !

Membership

Dongguan Astrand Electronic Technology Co., Ltd.
Free Member ·1YRS
,
Other

Main Products: 试验仪器

Xwboo>Products

Dongguan Astrand Electronic Technology Co., Ltd.Free·1YRS

  • E-mail

    sarah@astrand-martindale.com

  • Phone

  • Address

    No. 94, Chenwu Section, Guantai Road, Houjie Town, Dongguan City

Contact Now

XD-D03 2 Boxes High-Low Temperature Thermal Shock Material Testing Machine Two-box thermal shock ...

Model
Nature of the Manufacturer
Other
Product Category
Environmental test chamber
Place of Origin
City, Province
Overview
XD-D03 2 Boxes High-Low Temperature Thermal Shock Material Testing Machine Two-box thermal shock tester
Product Details

Product use:

temperature shock testing machine use: electronic and electrical components, metals, chemical materials, automation components, communication components, defense industry, aerospace industry, BGA, PCB substrates, electronic chips IC, semiconductor ceramics and physical properties of polymer materials Ideal test tool for changes. Ideal for physical changes in electronic and electrical components, metals, chemical materials, automation components, communication components, defense industry, aerospace industry, BGA, PCB substrates, electronic chips IC, semiconductor ceramics and polymer materials test tools.


Technical Features:

(1) With disk USB recording function, no recorder is required

(2) Flow calculation control of refrigerant servo valve. Save 30% of electricity

(3) High-temperature and high-humidity test to prevent condensation, dripping water technology industry first

(4) In the low temperature working state, the heater does not participate in the work, and the refrigerant flow and flow direction of the refrigeration unit are controlled and adjusted by PWM technology, and the three-way flow of the refrigeration pipeline, cold bypass pipeline and hot bypass pipeline is adjusted to realize the temperature control of the working chamber. Automatically constant. This method can reduce energy consumption by 40% under low temperature conditions

(5) Temperature synchronous rising and falling slope control (IEC68-2-56, 68-2-30), no dripping, no condensation and condensation

(6) Refrigeration advantages: traditional equipment low temperature control method: Ø Refrigeration compressor start and stop control temperature (large temperature fluctuation, seriously affect compressor life, obsolete technology)

Ø Constant operation of refrigeration compressor + heating PID control (causing the offset of cooling capacity and heating to realize temperature dynamic balance and waste a lot of electric energy)

Ø The new PWM cold control technology realizes low temperature energy-saving operation: the new PWM cold control technology realizes low temperature energy-saving operation


Technical parameter:

Preheating heating time

+20~+180℃, ≤60min

Pre-cooling and cooling time

+20~-55℃, ≤70min, +20~-65℃, ≤80min, +20~-75℃, ≤90min

Temperature conversion time

≤10 seconds

Overshoot of heating and cooling

≤±3℃

Temperature stabilization time

≤3MIN

Temperature fluctuation

±0.5℃

Temperature deviation

≦±2℃

Temperature uniformity

≦±2℃

Power

AC380V 50Hz

Test environment conditions

the ambient temperature is +28℃, the relative humidity is ≤85%, and there is no sample in the test box (unless otherwise stated)

Volume (80L) size

40x50x40(WxHxD)CM, external size is about 130x220x175(WxHxD)CM