Welcome Customer !

Membership

Dongguan Astrand Electronic Technology Co., Ltd.
Free Member ·1YRS
,
Other

Main Products: 试验仪器

Xwboo>Products

Dongguan Astrand Electronic Technology Co., Ltd.Free·1YRS

  • E-mail

    sarah@astrand-martindale.com

  • Phone

  • Address

    No. 94, Chenwu Section, Guantai Road, Houjie Town, Dongguan City

Contact Now

XD-D04 Lab UV Aging Chamber Industrial Quv Accelerated Weathering Tester UV aging tester

Model
Nature of the Manufacturer
Other
Product Category
Environmental test chamber
Place of Origin
City, Province
Overview
XD-D04 Lab UV Aging Chamber Industrial Quv Accelerated Weathering Tester UV aging tester
Product Details

Product Usage:

1. Ultraviolet light accelerated aging tester is mainly used to simulate the destructive effects of sunlight, humidity and temperature on materials;

2. Material aging includes fading, loss of light, strength reduction, cracking, peeling, chalking and oxidation, etc.;

3. The ultraviolet light aging test chamber can reproduce the damage that may occur outdoors for months or years by simulating sunlight, condensation, and natural humidity, and the sample is tested in a simulated environment for several days or weeks;

4. In the ultraviolet accelerated aging test machine, the fluorescent ultraviolet of the ultraviolet lamp can reproduce the influence of sunlight, and the condensation and water spray system can reproduce the influence of rain and dew;

5. The temperature is controllable throughout the test cycle;

6. A typical test cycle is usually UV exposure at high temperature and a dark humid condensation cycle at 99% relative humidity;

7. Typical applications in paint coatings, automotive industry, plastic products, wood products, glue, anodized film, etc.


Test conditions:

1. The test sample is fixed on the test rack, facing the fluorescent lamp. When the sample is not filled with the sample rack, the blackboard should be used to fill the sample rack to keep the inner wall of the test box closed.

2. Test temperature. Three temperatures of 50, 60, and 70°C can be used for illumination, and 60°C is preferred; the temperature in the condensation stage is 50°C, and the temperature tolerance uniformity is ±3°C.

3. There are two cycles of 4H illumination, 4H condensation or 8H illumination and 4H condensation for each condensation cycle of illumination.

4. After the first illumination of 400-450H, each row of lamps needs to be replaced with a fluorescent lamp, and other lamps shall compensate for the loss caused by lamp aging according to the lamp cycle and replacement method. This operating mode ensures that the service life of each lamp is 1600-1800H.

5. When replacing the lamp, the water tray should be wiped dry and cleaned to avoid the formation of limescale.


Technical parameter:

Temperature range

RT+10℃~70℃

Temperature uniformity

±3℃ Temperature fluctuation: ±0.5℃

Humidity range

≥75%RH

The distance between the center of the lamp tube

65~70mm

The distance between the sample and the center of the lamp

55mm

Ultraviolet tube

UV-A340, its peak value is 340nm, and the normal service life of American Atlas/Q-LAB is more than 2000 hours.

Standard test piece

75×150mm

Irradiance intensity

light irradiance controllable 0.35-1.1W/㎡ / adjustable 340nm band (irradiance intensity value set through touch screen)

Irradiance control

intelligent control of irradiance, through real-time feedback of irradiance, automatic compensation of the set value (optional). (If there is no irradiation probe, remove this one)

Spray system

spray water for 18 minutes, stop spraying for 102 minutes, the time can be set

Studio size

450×1165×500mm (D×W×H)

Dimensions

D600XW1325XH1480mm

Liner material

SUS304 stainless steel plate

Shell material

SUS304 stainless steel plate

Lamp tube

UVA-340 ultraviolet lamp tube, a total of 8 in the second row, 40W / only (a total of 8)

Water source and consumption

clean water or distilled water 8 liters/day (customers bring their own)