XD-A21 Laboratory Paper Air Permeability Tester (Schopper)

Model
Nature of the Manufacturer
Other
Product Category
Paper and Paperboard Testing Equipment
Place of Origin
City, Province
Overview
XD-A21 Laboratory Paper Air Permeability Tester (Schopper)
Product Details

Application: 

The paper air permeability tester is designed and produced according to the national standard QB/T1667 "Paper Air Permeability (Schopper Method) Tester". This instrument is specially used for the performance testing of material air permeability. It can be applied to many kinds of paper, such as cement bag paper, paper bag paper, cable paper, copy paper and industrial filter paper, etc. It is necessary to measure the degree of air permeability.

This paper air permeability tester is suitable for paper with air permeability between 1×10-2-1×102um/(Pa.S), not suitable for paper with large surface roughness.


Standard:

ISO 5636.1, GB/T 458, QB/T1667


Parameter:

Measuring rangeA range (0~1000)ml/min, B range (0~5000)ml/min, ~10000)ml/min, D range (0~100000)ml/min(B, C, D gears are optional gears)
Test area

The standard sample test is 10±0.2cm², and the sample must be larger than this size.

Test area pressure differenceThe standard test pressure difference is 1±0.01kpa, and other pressure differences can be customized for special materials.
Measuring head304# stainless steel air permeability measuring fixture, sealed with anti-aging silicone.
Measurement accuracy<100ml, error="" 1ml;="">100ml, error 5ml
The material of the upper and lower partitions of the graduated cylindercorrosion-resistant material, steel, to prevent corrosion problems after long-term use
Barometer BracketAnti-corrosion construction using anodized aluminum alloy
Catheter materialSilicone catheter is used to eliminate the aging problem caused by long-term use
Valve materialuse pure copper valve to prevent the phenomenon of rusting and blocking the valve
Inner diameter of clamping ringstandard sample clamping inner diameter 35.68±0.05mm
Coaxiality of the central hole of the upper and lower clamping ringsstrictly control the coaxiality, the deviation is less than 0.05mm
Ambient conditions Temperature(0~35) ℃, humidity <>
Dimensions400*300*1200mm
WeightAbout 38kg