KN-482 Ash Content for Petroleum Product

Model
Nature of the Manufacturer
Other
Product Category
Place of Origin
City, Province
Overview
KN-482 Ash Content for Petroleum Product
Product Details

ash.png

KN-482 Ash Content for Petroleum Product

Overview

    This oven conforms to the ASTM D482 Standard Test Method for Ash from Petroleum Products and ASTM D874 Standard Test Method for Sulfated Ash from Lubricating Oils and Additives it is used for determination of ash in the range 0.001–0.180 mass %, from distillate and residual fuels, gas turbine fuels, crude oils, lubricating oils, waxes, and other petroleum products, in which any ash-forming materials present are normally considered to be undesirable impurities or contaminants. The apparatus is limited to petroleum products which are free from added ash-forming additives, including certain phosphorus compounds.

 

Features

1. The inner furnace adopts sealed structure, and we adopts high temperature resistant lightweight refractory for the furnace mouth and furnace door.

2.The insulating layer between the inner furnace and the furnace shell adopts refractory fiber, expanded perlite products, there is trace amounts of heat on the outer shell, even if the temperature in the furnace rises to 775℃;

3. The furnace door adopts clip-on structure, only one action to close or open, conforms to quick take of the ASTM D482 standard;

 

Technical parameters

1. Rated voltage: AC220V±10% 50Hz

2. Heating power: 5000W

3. Temperature control mode: temperature controller

4. Temperature measurement mode: nickel-chromium——Nickel-silicon galvanic couple

5.Temperature control precision:±5℃

6.Temperature control range:100~1200℃

7.Ambient requirements:Temperature:5~40℃

8. Humidity≤85%