KN-6481 Portable Energy Dissipation X-ray Fluorescence Spectrometer

KN-6481 Portable Energy Dissipation X-ray Fluorescence Spectrometer
Overview
KN-6481 Portable Energy Dissipation X-ray Fluorescence Spectrometer conforms to ASTM D6481 Standard Test Method for Determination of Phosphorus, Sulfur, Calcium, and Zinc in Lubrication Oils by Energy Dispersive X-ray Fluorescence Spectroscopy and ASTM D7751 Standard Test Method for Determination of Additive Elements in Lubricating Oils by EDXRF Analysis. This tester is used to test metal elements in oils like P, S, Cl, Ca, Ti, V, Cr, Mn, Fe, Ni, Cu Zn, PB, Mo, Ag, Cd, Sn, etc.
Features
1. Portable and one button operation.
Working Curve Linear Chart

Sample Spectrogram Comparison

Repeatability Data
Fuel Oil 50ppm Sulfur, Lubricating Oil 30ppm Heavy Metal Stability Data |
Repeatability | S (PPM) | Ti (PPM) | V (PPM) | Mn (PPM) | Ni (PPm) | Zn (PPM) | Pb (PPM) |
1 | 49.52 | 30.71 | 29.58 | 30.11 | 29.54 | 29.74 | 31.22 |
2 | 48.68 | 31.05 | 30.14 | 31.25 | 28.76 | 30.52 | 30.41 |
3 | 49.33 | 31.19 | 28.96 | 31.10 | 29.35 | 31.21 | 30.63 |
4 | 50.12 | 31.29 | 30.31 | 31.51 | 28.41 | 29.65 | 30.33 |
5 | 48.76 | 31.42 | 30.11 | 29.65 | 27.96 | 28.55 | 29.42 |
6 | 49.62 | 31.07 | 29.75 | 29.51 | 29.34 | 29.41 | 28.53 |
7 | 49.53 | 30.75 | 29.51 | 29.41 | 30.04 | 30.66 | 29.58 |
8 | 50.18 | 30.54 | 29.41 | 30.65 | 27.53 | 29.64 | 30.11 |
9 | 51.83 | 30.62 | 28.96 | 30.74 | 28.55 | 29.34 | 29.44 |
10 | 49.72 | 30.88 | 29.47 | 30.88 | 29.43 | 30.63 | 30.98 |
11 | 49.33 | 30.08 | 30.11 | 31.54 | 29.01 | 29.75 | 28.87 |
Average Value | 49.69 | 30.87 | 29.66 | 30.58 | 29.90 | 29.83 | 29.96 |
Standard Deviation | 0.85 | 0.39 | 0.47 | 0.79 | 0.74 | 0.85 | 0.86 |
RSD (%) | 1.71 | 1.25 | 1.57 | 2.58 | 2.58 | 2.85 | 2.87 |
Technical Parameters
1. Analysis method: Energy Dissipation X-ray Fluorescence
2. Detector: Fast SDD, resolution can reach 125eV
3. Excitation source: 50KV/100uA – silver target end window integrated micro X-ray tube and high voltage power supply
4. Elements measuring range: All elements from Al (Aluminum) to U (Uranium)
5. Detection time: 60~200s
6. Sample type: Solid, Liquid, Powder
7. Detection limit: Fuel oil: P 3ppm, S 1ppm, Cl 2ppm
Lubricating oil: Ti 0.5ppm, V 0.5ppm, Cr 0.5ppm, Mn 0.3ppm, Fe 0.8ppm, Ni 1ppm, Cu 0.5ppm, Zn 0.5ppm, Pb 0.6ppm, Mo 1ppm, Ag 6ppm, Cd 8ppm, Sn 10ppm
8. Content range: 5ppm~99.99%
9. Calibration method: Cu
10. Processor (CPU): 1G Hz
11. Operation: One button operation, no need to select the corresponding test mode
12. Relative humidity≤90%
13. Operation temperature range: -20~50℃
Standard configuration
1. Fast SDD Multichannel Detector
2. X-ray tube, high voltage and low voltage power supply
3. Amplifying circuit and digital multi-channel system
4. Power and control system
5. Built-in PDA
6. Dedicated software, PDA version
7. Bluetooth printer (Optional)
8. 110V/220V Universal adapter
9. Shock-resistant, pressure-resistant, lockable carrying case