KN-170Z Automatic Abel Closed Cup Flash Point Tester

Model
Nature of the Manufacturer
Other
Product Category
Place of Origin
City, Province
Overview
KN-170Z Automatic Abel Closed Cup Flash Point Tester
Product Details

abel.png

KN-170Z Automatic Abel Closed Cup Flash Point Tester

Overview

KN-170Z Automatic Abel Closed Cup Flash Point Tester conforms to IP170 Petroleum products and other liquids -- Determination of flash point -- Abel closed cup method and ISO 13736 Petroleum Products and Other Liquids - Determination of Flash Point - Abel Closed Cup Method. This International Standard specifies a method for the determination of the closed cup flash point of petroleum products and other liquids having flash points between -30℃ and 70℃ inclusive.   However, the precision given for the method is only valid for flash points in the range -5℃ to 66.5℃. This standard is not acceptable to water borne paints which may however be tested using ISO 3679


Features

1. Dual control system: The tester adopts the hardware system of industrial computer and single-chip microcomputer. Uses the single-chip microcomputer to control the heating process and the data acquisition, so that the test is high in real time and the test results are more accurate. The industrial computer as the overall system control makes the tester more reliable

2. Multiple ports: RS-232, USB and RJ-45, etc. Be able to connect to the LIMS system, LAN, surf the internet. Be able to do update via the USB port and also can print the test results by connecting to the laser printer

3. Color touch screen operation, also equipped with mouse and keyboard

4. Double Ignition method: Electric Ignition or gas Ignition, both of them have no effect to the results

5. Gold plating parts: Sample cup and lid are all gold plated, for a great performance

6. Imported temperature sensor from Germany

7. Self detection, if there is any part is abnormal, the tester will buzz and alarm

8. The tester is able to display the heating rate in graph and figure

9. Over temperature protection, if the test temperature exceeds the safe temperature, the tester will stop the test automatically and start to do cooling

10. Atmospheric pressure correction function, and be able to query the result as exact demands.


Technical parameters

1. Test temperature range: -30~100℃ (External circulating cooling water bath)

2. Temperature sensor: Platinum resistance PT100

3. Display mode: 6.4 inch TFT LCD

4. Cooling mode: Semi conductor

5. Heating rate: 1℃ / min

6. Temperature accuracy: 0.1℃

7. Stirring speed: 0.5 r/s (30r / min)

8. Ignition mode: Electric or gas ignition

9. Main host power consumption: ≤450W, Circulating bath≤1200W

10. Rated voltage: AC220V±10%, 50Hz

11. Weight: 25kg