KN-93E Automatic Pensky-Martens Closed Cup Flash Point Tester

Model
Nature of the Manufacturer
Other
Product Category
Place of Origin
City, Province
Overview
KN-93E Automatic Pensky-Martens Closed Cup Flash Point Tester
Product Details

93E.png

KN-93E Automatic Pensky-Martens Closed Cup Flash Point Tester

Overview

KN-93E Automatic Pensky-Martens Closed Cup Flash Point Tester conforms to ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to test for those liquids which have a kinematic viscosity of 5.8mm2/s (cSt) or more at 37.8℃ or 9.5mm2/s (cSt) or more at 25℃ (77℉), or that contain suspended solids, or have a tendency to form a surface film while under test. These departments for liquids using these test methods have established other classification flash points.


Features

1.        Adopts color touch LCD screen

2.        Automatic heating control

3.        Automatic open the cup cover

4.        Automatic ignition and cooling

5.        Automatic atmospheric pressure correction

6.        After detecting the flash point, the tester will display, print and save the results automatically


Technical parameters

1.      Measuring range: Ambient~250℃

2.        Accuracy

          Flash point≤110℃, ±2℃

          Flash point>110℃, ±4℃

3.        Detection mode: Thermocouple differential detection

4.        Ignition mode: Automatic electronic ignition

5.        Screen: 7 inch touch screen

6.        Cooling mode: Fan cooling

7.        Working temperature: 10~35℃

8.        Related humidity: 30~80%

9.        Rated voltage: AC220V±10%, 50Hz±5%

10.    Power≤450W