1. LBC- ⅱ A2 constant air flow 99% discharge, negative pressure vertical laminar flow.2. Moving glass door up and down, can be arbitrarily positioned, easy to operate, and can be completely closed for sterilization, positioning height limit alarm prompt.3. Equip the power output socket of the working area.4. Hepa filter is suitable for filtering particles, smoke and microorganisms in the air. Two layers of HEPA filter for 99.999% or more.5. The working environment adopts high quality 304 stainless steel, smooth, s......
Scope of application:
Biosafety cabinet is a necessary equipment in modern microbiology laboratory, especially for those occasions where operators need to take protective measures.
Such as medical, pharmaceutical, microbiology, biomedicine, gene recombination, animal testing and biological products,
Bacteria culture to provide both bacteria - free dust - free and safe working environment.
Product type
| LBC-1600IIA2 |
Clean level | ISO 5(100级Class 100)/ISO 4(10级Class 10) |
Filter level Filter efficiency | @0.3~0.12µm≥99.999% ~ 99.995% |
Mean downdraft velocity | 0.25m/s~0.5m/s |
Average flow velocity of incoming air | ≥0.5m/s |
Noise | ≤60dB(A) |
Half peak of vibration | ≤5µm |
Power | AC,220V/50Hz |
Product protection (repeat 3 times) | Total number of colonies≤5CFU/time |
Cross-contamination protection (repeat 3 times) | Total number of colonies≤2CFU/time |
Maximum power | ≤1500VA(with spare socket) |
Weight(kg) | 250 |
Size(W×L×H)(mm) | 1600×570×635 |
Surface size(mm) | 1800×795×2350 |
fluorescent lamps Specification and quantity | 40W×② |
Uv lamp specifications and quantity | 40W×① |
Intensity of illumination | ≥650Lx |
Exhaust pipe diameter | Φ250 |
Wind direction | A2 |