UV-1300 UV-visible spectrophotometer is a high-quality scanning equipment, through the computer online, easily scan the whole band, and can complete the dynamics analysis, quantitative analysis and multi-wavelength analysis, is an extremely broad range of applications practical apparatus.
Features
Single-chip microcomputer control, LCD Screen (128*64 Dots)
Large LCD display can display multiple sets of data.
Large memory space can store multiple sets of data and curve.
Auto zero,99% adjustment function
Auto wavelength adjustment
Auto filter changing, large sample pool (5mm ~ 100mm)
With up to ten standard sample for measurement function of standard curve establish
Direct input by K, B factor for quantitative measurement standard curve
Can be directly input and the corresponding standard concentration values establish a standard curve for quantitative measurement
Switch off power save standard curve parameter measuring set
With general purpose parallel printer port, you can print the title bar, the measurement data, curve parameters, standard sample point and the curve
With USB interface
Can achieve more accurate and flexible measurement requirements (optional) via the PC control Technical Specifications
Application
Drug analysis: National Formulary" Available UV-visible spectrophotometry drugs for drug testing, drug analysis, pharmaceutical and other industries
Life Sciences: DNA, protein concentration
Animal husbandry and fishery: Pesticide residues, crop composition testing, veterinary testing, feed testing, fertilizer testing, soil composition testing, testing and other aquaculture
Metal elements:Ag Cu Pb Zn Co Ni K Na Ca Mg Li Mn Cr As, including precious metals
Geological Survey: Determination of metal elements in minerals and inorganic salts
Food safety: Content of additives, preservatives, perfumes, fats, enzymes, sugars, spices, minerals, vitamins, etc.
Environmental monitoring:Monitoring water, air, rainfall and soil, etc., in which the determination of various pollutants
Wavelength Range | 190~1100nm |
Band Width | 4nm |
Wavelength Accuracy | ±0.5nm |
Wavelength Repeatability | ≤0.2nm |
Photometric Accuracy | ±0.3% τ |
Photometric Repeatability | 0.15% τ |
Stray Light | ≤0.05% τ (220nm NaI,340nm NaNO2) |
Stability | 0.001A/h@500nm |
Photometric Mode | T,A,C,E |
Wavelength Setting | Automatic |
Photometric Display Range | -0.3~3A |
Display Mode | LCD Screen (128*64 Dots) |
Detector | Import Silicon Photodiode |
Light Source | Import Deuterium Lamp & Tungsten Halogen Lamp |
Power Requirement | AC 220V/50Hz 110V/60Hz |
Power | 120W |
Dimensions(W*D*H) | 480×350×220mm |
Weight | 15Kg |