Model | YS6020(Pulsed xenon lamp) |
Optical Geometry | Reflect: di:8°, de:8°(diffused illumination, 8-degree viewing angle); SCI (specular component included)/SCE (specular component excluded);Include UV / excluded UV light source; Transmittance: di:0°, de:0° (diffuse illumination: 0° viewing); SCI (specular component included)/SCE (specular component excluded);Include UV / excluded UV light source; Haze(ASTM D1003); Conforms to CIE 5,GB/T 3978,GB 2893,GB/T 18833, ISO7724/1, ASTM E1164, DIN5033 Teil7 |
|
|
Integrating Sphere Size | Φ 154mm |
Light Source Device | 360nm-780nm Xenon Lamp, 400nm cut-off Xenon Lamp |
Spectrophotometric Mode | Concave Grating |
Sensor | 256 Image Element Double Array CMOS Image Sensor |
Wavelength Range | 360-780nm |
Wavelength Pitch | 10nm |
Semiband Width | 5nm |
Reflectance Range | 0~200% |
Measuring Aperture | Reflective: Φ30mm/Φ25.4mm, Φ10mm/Φ8mm, Φ6mm/Φ4mm; Transmissive: Φ25.4mm; Remark: 1. Automatic identification of switch caliber 2. Customized Configuration caliber and lens position |
|
Specular Component | Reflectance: SCI&SCE Transmittance: SCI&SCE |
|
Color Space | CIE Lab, XYZ, Yxy, LCh, CIE LUV, Hunter LAB, Munsell, s-RGB, HunterLab, DIN, βxy |
Color Difference Formula | ΔE*ab, ΔE*uv, ΔE*94, ΔE*cmc(2:1), ΔE*cmc(1:1), ΔE*00v, ΔE(Hunter), DIN ΔE99 |
Colorimetric Index | WI (ASTM E313, CIE/ISO, AATCC, Hunter), YI (ASTM D1925, ASTM 313), MI (Metamerism Index), Staining Fastness, Color Fastness, Color Strength, Opacity, Gardner Index, APHA/Pt-Co Index, 555 Index, Haze(ASTM D1003) |
|
|
|
|
|
Observer Angle | 2° & 10° |
Illuminants | D65,A,C,D50,D55,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,DLF,TL83,TL84,TPL5,U30 |
Displayed Data | Spectrogram/Values, Chromaticity Values, Color Difference Values/Graph, Pass/Fail Result, Color Offset |
Measuring Time | About 2.4s (Measure SCI & SCE about 5s) |
Repeatability | Spectral reflectance: Φ25.4mm/SCI, Standard deviation within 0.07% Chromaticity value: Φ25.4mm/SCI, Standard deviation within ΔE*ab 0.015 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration) Spectral Transmittance: Φ25.4mm/SCI, Standard deviation within 0.07% Chromaticity value: Φ25.4mm/SCI, Standard deviation within ΔE*ab 0.018 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
|
|
|
Inter-instrument Error | Φ25.4mm/SCI, Within ΔE*ab 0.15 (Average for 12 BCRA Series II color tiles) |
|
Size | 370×300×200mm |
Weight | About 9.6kg |
Power Supply | DC 24V, 3A Power Adapter |
Light Source Device Life | 5 years, more than 3 million times measurements. |
Display | 7" TFT Capacitive Screen-touch Display |
Data Port | USB & Print serial port |
Data Storage | Standard 2000 Pcs, Sample 20000 Pcs |
Language | Simplified Chinese, Traditional Chinese, English, |
Working Environment | Temperature: 0~40℃; Humidity: 0~85% (No Condensation) |
Storage Environment | Temperature: -20~50℃; Humidity: 0~85% (No Condensation) |
Standard Accessory | Power Adapter, User Guide, CD Disk(PC Software), USB cable, Standard Calibration Board, Black Calibration Cavity, Transmission black baffle, Sample holder, 25.4mm caliber, 8mm caliber, 4mm caliber, |
Optional Accessory | Micro-printer, Transmissive Test Clamp Component,Instrument inversion components,culture dish |
Notes: The specifications are subject to change without notice. |