Description:CPA-3A type portable laser respirable dust continuous tester for the measurement of respirable dust in air of public places (PM10) concentration and number of clean environment in unit volume of air particle size of 0.5 - 10 m dust particles. It is a
CPA-3A type portable laser respirable dust continuous tester for the measurement of respirable dust in air of public places (PM10) concentration and number of clean environment in unit volume of air particle size of 0.5 - 10 m dust particles. It is a new generation of intelligent measuring instrument, which is controlled by the Centers for Disease Control and prevention, health supervision and environmental monitoring.
The instrument uses advanced light scattering principle to the semiconductor laser as the measurement light source, the high sensitive non-contact measurement of dust in the air. The use of high-resolution large screen high brightness LCD display, all Chinese display.
The utility model has the advantages of small size, convenient carrying, high testing speed, high sensitivity, good stability, simple operation, etc.. Suitable for measuring without external power supply. The utility model can be widely used in various public places and electronic, optical, food, aerospace and
The apparatus according to China's health industry standard: "public places of inhalable particulate matter (PM10) method for the determination of light scattering method" (WS/T 206-2001) design.
CPA-3A (single test PM10)
1、PM10 (particle size less than 10µm m respirable dust particles) measurement range: 0.001~10mg/m fand (GB)
2、measuring particle size: the smallest particle size is greater than 0.3um, the maximum particle size is less than 10um
3、detection resolution: 0.001mg/m fand;
4、sampling time: 60s;
5、data storage capacity: 200 sets of data; with the computer: network transmission can be achieved. Optional printer;
6、working environment: temperature humidity (0-39 ℃) RH (2070%);
7、power supply: 220V AC power supply; built in high-capacity lithium battery, can be used continuously for more than 4 hours after adequate electricity;
8、volume: 110mm * 220 mm * 50 mm; weight: about 580g.
Standard configuration
(1) a hosts
(2) a USB-RS232 interface program CD
(3) a dedicated USB-RS232 cable
(4) a power charger
(5) a Aluminum Alloy box
(6) a sets of external temperature and humidity senso
Optional: micro printer, charger