Welcome Customer !

Membership

Shandong Drick Instruments Co.,LTD
Free Member ·1YRS
Other

Main Products: 试验机

Xwboo>Products

Shandong Drick Instruments Co.,LTDFree·1YRS

  • E-mail

    info@drktest.com

  • Phone

    0086-531-85868997,0086-531-85761369

  • Address

    3rd Building, A District,Shidai headquarters, No. 15 Lanxiang Road, Tianqiao District,Jinan City,Shandong Province, China. 250032

Contact Now

DRK710A Melt Flow Updater GB/T9643, GB/T3682.1, JB/T5456, ISO1133

Model
Nature of the Manufacturer
Other
Product Category
Plastic soft package testing instrument
Place of Origin
City, Province
Overview
DRK710A Melt Flow Updater GB/T9643, GB/T3682.1, JB/T5456, ISO1133
Product Details

Product details

Features

1. Using PID intelligent temperature control, the temperature control accuracy is high, which can ensure that the temperature at the top 10 mm of the die in the barrel is within ±0.5℃

2. The key parts are nitrided, with high strength, hardness and small deformation characteristics, and ensure the accuracy and stability of the measurement data;

3. The heating speed is fast. The heater of the instrument adopts a special spring heating ring for hot runner. Its main material is imported nickel-chromium wire alloy, which is better than traditional dynamic heaters.

4. Stronger dry burning resistance, higher power, better heating uniformity and long service life, supporting the instrument to work for a long time at the highest temperature of 450 ℃;

5. The instrument heating adopts a voltage regulating module, which can adjust the voltage and power in a full range. Compared with the traditional solid-state relay on-off control, the heating is more stable and rapid, so that the constant temperature state can be quickly restored after filling;

6. The instrument adopts a customized armored platinum resistance temperature sensor with the highest accuracy class A, and the accuracy is within 0.1℃. Compared with traditional sensors, it has the characteristics of impact resistance, corrosion resistance, vibration resistance, and can be bent at will. Longer life, can ensure the long-term use of the instrument under the highest temperature of 450 ℃ environment.

Technical parameter

1. Constant temperature accuracy: ±0.5℃

2. Temperature range: 50-450℃ 3. Temperature recovery time: 2min

5. Timing accuracy: 0.1S

6. Test method: MFR quality method, MVR volume method

7. Test range: 0.1-80g/10min (MFR), 0.1-3000g/10min (MVR)

8. Maximum cutting times: 999 times

9. Maximum cut-off time: 999 seconds

10. Piston rod head diameter: Φ9.475±0.015mm